3. Booting issue trouble-shooting
3.1 Connect USB cable in the another port of the computer.
3.2 Check to see if another USB is in working order by swapping with a working one.
4.3 Remove the top cover and internal cover.
4.4 Check the core board led status
4.5 When you turn on the power, one led of the core board is on.
(One led is on → Three led is on → One led is on.)
4.6 If the led of the core board is something wrong, replace the core board.
4.7 Check the power board
4. Not recognize the cassette housing
4.1 Scan it with another cassette housing to check cassette working well.
4.2 Check the RFID board.
- Refer to the Service Manual p.32 for RFID board replacement.
Summary of Contents for FireCR
Page 1: ...DIGIRAY CORP Basic Guide of FireCR Scanner...
Page 5: ...for User manual download for Quantor software download...
Page 10: ...the Windows Security Dialog Box pops up after installation click Allow access...
Page 11: ...7 PC requirement Recommended PC requirement Minimum PC requirement...
Page 13: ...Basic configuration of CR Computed Radiography system Imaging Plate Cassette Housing...
Page 16: ...FireCR Scanner Video on how to use https www youtube com watch v xVkq_ sIabc...
Page 20: ...Main Board Core Board...
Page 33: ...Understanding FireCR Flash Calibration Note X ray exposure must cover entire cassette...
Page 41: ...2 Disassembly Top Cover 1 Pull back cover...
Page 48: ...2 Disassembly BLDC Board 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order...
Page 49: ...2 Disassembly Main Board 1 Unscrew the 6 screws that hold down the board and replace 2 5mm hex...
Page 54: ...2 Disassembly Aligner Motor 1 Remover back over...
Page 61: ...2 Disassembly 5 Remove the two screws at the edge of the eraser 6 Pull back eraer Fiber Bundle...
Page 74: ......