2. Disassembly
Internal Cover
3. Lift off internal Cover
Note: There are two grooves on each side of the Internal Cover front. When refitting the Internal Cover,
make sure that the cables on both sides go through the upper grooves. Refit it from front to back and
make sure that the end of the Internal Cover goes over the entire internal frame and not into the space between
the internal frame and the back cover.