Main Board & Core Board
• Function
1. Control all peripherals of for acquisition of image data using FPGA and MCU.
- MCU and FPGA always communicate each other.
- Sensor reading status is transferred to the FPGA and to MCU.
- MCU and FPGA give a command to some peripherals such as Y-axis motor and Eraser unit etc.
2. All sensors on/off reading.
3. Convert analog signal to digital.
4. Provide Power to PMT, laser module, RFID board & some peripherals..
5. Transmit image data to PC using USB communication.
• Main board Issue
1. Stop during mid-scan.
– main board
2. Wavy noise.
– main board
3. ScanLow value is abnormal.(White image)
– main board
4. Partial horizontal lines on image.
– main board
5. Booting fail.
– core board
Summary of Contents for FireCR
Page 1: ...DIGIRAY CORP Basic Guide of FireCR Scanner...
Page 5: ...for User manual download for Quantor software download...
Page 10: ...the Windows Security Dialog Box pops up after installation click Allow access...
Page 11: ...7 PC requirement Recommended PC requirement Minimum PC requirement...
Page 13: ...Basic configuration of CR Computed Radiography system Imaging Plate Cassette Housing...
Page 16: ...FireCR Scanner Video on how to use https www youtube com watch v xVkq_ sIabc...
Page 20: ...Main Board Core Board...
Page 33: ...Understanding FireCR Flash Calibration Note X ray exposure must cover entire cassette...
Page 41: ...2 Disassembly Top Cover 1 Pull back cover...
Page 48: ...2 Disassembly BLDC Board 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order...
Page 49: ...2 Disassembly Main Board 1 Unscrew the 6 screws that hold down the board and replace 2 5mm hex...
Page 54: ...2 Disassembly Aligner Motor 1 Remover back over...
Page 61: ...2 Disassembly 5 Remove the two screws at the edge of the eraser 6 Pull back eraer Fiber Bundle...
Page 74: ......