TEST AND AJUSTMENT MANUAL
49-TPOSE-PSC-EN02
9
4.
ADJUSTMENT
These functions allow the user to program, adjust and calibrate the scale. IMPORTANT NOTE: this should
only be done by authorized personnel.
The scale must be turned on for at least two hours before being adjusted. It must also be perfectly level and
the temperature and humidity must be stable. The user must have a calibrated weight relative to the scale
capacity.
Switch on the scales and enter test and adjustment mode. To do this, during countdown from 9 to 0, press
and
at the same time. The scales will go into test mode. Then press the adjustment key situated on
top of the CPU. This key can be pressed by removing a security-covered screw and inserting a pointed
object in the hole.
The following may be programmed:
4.1. COUNTRY PROGRAMMING.
This parameter is only used for consultation. Change of country must be remote loaded in the scale.
Press
to go to the next parameter.
4.2. OIML PROGRAMMING.
This allows the user to set a series of parameters as recommended by the OIML. To do so, program a 1 as
the default value. This will also appear each time the country programmed is changed.
Press
to go to the next parameter.
4.3. CAPACITY PROGRAMMING.
This allows the scale capacity to be selected according to the following table.
The following is the capacity table.
Code
Maximum Capacity
1
6 kg
2
15 kg
3
30kg
16
6/15kg MR
17
15/30 kg MR
Press
to reduce the value and
to increase it.
Press
to go to the next parameter.