![Diamond SmartLyte Service Manual Download Page 60](http://html.mh-extra.com/html/diamond/smartlyte/smartlyte_service-manual_2493455060.webp)
5 Maintenance and Troubleshooting
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 60 of 81
VALVE OVERTEMP! INSPECT VALVES
In case one of the valve solenoids overheats or is not connected properly, the message
VALVE OVERTEMP! INSPECT VALVES will be displayed. Each solenoid
incorporates a thermo-fuse, which will reset automatically.
Possible causes and remedies:
•
Turn power off for at least one hour.
•
Check fuse F3 (24V).
•
Check whether room temperature is within specified limits (15°C to 32°C / 60°F to 90°F).
•
Ensure all the pinch valves are connected properly.
Summary of Contents for SmartLyte
Page 1: ......
Page 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Page 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Page 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Page 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Page 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Page 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Page 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...