3 Software Operation
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 31 of 81
ANALYZER RESET
Press
YES
and
NO
keys simultaneously and turn power off and back on to provide a total reset
of the instrument.
CAUTION!
All QC Data and Instrument programming will be lost.
Switching from human to vet or vet to human resets all previously saved data as well.
FAIL SAFE SQ UPLOAD
(SW Versions 12.07.C5 and 13.07.C5 and higher)
In situations where instrument doesn’t initialize completely due to wrong software upload or
peripheral issues, bootloader mode which allows upload of software to clear the error can be
entered directly by following these steps –
1.
Turn instrument off.
2.
Press YES button.
3.
While pressing the YES button, turn on the instrument.
4.
Wait for 30 seconds.
5.
Instrument is now in programming mode.
6.
New software can be uploaded using provided steps.
WARM-UP BYPASS
Press
YES
and
NO
keys simultaneously during instrument warm-up.
Summary of Contents for SmartLyte
Page 1: ......
Page 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Page 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Page 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Page 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Page 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Page 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Page 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...