3 Software Operation
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 30 of 81
ENTER CONFIG CODE
CONFIG CODES are available and allow additional programming features. Following is a
listing of the available CONFIG CODES and functions.
SET
RESET
Code Description
AME
EMA
Dialysates in Acetate matrix can be analyzed.
BCE
ECB
Dialysates in Bicarbonate matric can be analyzed.
CPC
Connect to PC to load software
CSC
Clears all codes and return to factor default
ECM
MCE
Allows the operator to run the instrument in the economy mode. The
instrument will enter the STANDBY MODE. This occurs at the time of
the next 2-point calibration when no samples have been analyzed since
the last 2-point calibration.
HRS
SRH
Provides one additional digit of resolution for each parameter for blood
and serum samples. QC and Standard samples are always displayed with
high resolution. Urine samples are always displayed in low resolution.
LCR
Prints lease counter report.
MGC
CGM
This config code is available only in the mode SMARTLYTE analyzer
and enables Ca++ values to be displayed in mg/dL. The normal mode
displays values in mmol/L. QC and Standard results are always
displayed in mmol/L.
PMI
This code will print the model information of the SMARTLYTE
PSC
Prints service codes.
QCF
FCQ
When this code is enabled, QC values will be reported in direct ISE
measurement instead of flame-equivalent values. In this mode,
correlation factors are NOT applied to the measured values.
TBO
OBT
This disables the beep.
TDV
VDT
Allows printing of the electrode voltages for sample measurements and
calibrations.
R14
V48
These codes allow the dealer to enter their own password for those
instruments will be released based on number of tests.
RST
Erases all data from memory including correlation factors, QC Ranges,
stored patient, QC data and lease count.
FPC
Displays the fluid pack counter used to monitor pack consumption.
Note:
For SW versions 12.07.C5 and 13.07.C5 and higher.
Summary of Contents for SmartLyte
Page 1: ......
Page 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Page 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Page 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Page 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Page 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Page 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Page 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...