5 Maintenance and Troubleshooting
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 58 of 81
ELECTRODE VOLTAGE RANGES
Allowed difference
Electrode Standard A Standard B Standard C A-B A-C
Na
+
- 600 to +2400 -1600 to +2000 -600 to +2400 +250 to +680 -50 to +50
K
+
- 700 to +1000 -2500 to +500 -700 to +1000 +470 to +1200 -40 to +40
Cl
-
- 3100 to +500 -1000 to +3000 -3100 to -100 -370 to -860 not used
Ca
++
- 3100 to +1000 -2300 to +2500 -3100 to +1000 -350 to -660 -150 to +150
Li
+
- 3100 to +1900 -3600 to +1400 -2600 to +3400 +1 to +760 -1730 to -285
INSPECT FLUID PATH
After completion of sample measurement or calibration, the analyzer will monitor the
sample sensor to determine if the sample path has been cleared. This message will be
displayed if the sample path has not been cleared.
Possible causes and remedies:
•
Check for blockages in the sample path, especially in the sample probe, the tubing to the
sample sensor and in the sample sensor.
•
Check sample sensor is securely plugged in and check sample sensor under SERVICE MENU
to ensure correct operation.
•
Check peristaltic pump and tubing are correct and all tubing is properly in place.
Run Prime Function, Run Calibration
In case the analyzer displays arrows up or arrows down instead of sample results, the
concentration of the sample is outside of the measurement range.
Possible causes and remedies:
•
If the sample is a urine sample, arrows up instead of the K+ result indicate that further dilution
of the specimen is required.
•
Check for proper sample preparation.
•
Ensure that the sample is correctly aspirated into the measuring chamber and ensure small air
bubbles are not present.
•
Check for proper aspiration of STD A.
Summary of Contents for SmartLyte
Page 1: ......
Page 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Page 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Page 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Page 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Page 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Page 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Page 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...