5 Maintenance and Troubleshooting
ECO 9085 SOP05-5038F Rev.05 Effective Date: 12/06/18 Page 57 of 81
INSPECT ELECTRODES
This message is displayed when either of the following conditions is present.
1.
More than six aspirations of STD A does not result in three consistent STD A readings for all
electrodes.
2.
The difference between voltage A and B is out of specification for all channels.
Possible causes and remedies:
•
Electrodes are not plugged in. Check the slide mechanism to ensure it is in the correct
position and firmly pushed into electrode connectors.
•
Perform MAINTENANCE.
•
Check or replace reference electrode.
•
Replace Fluid Pack.
•
Check connection from shielding to ground to determine if electrical short is present.
•
Call Product Support.
NA INVALID CAL
K INVALID CAL
CL INVALID CAL
CA INVALID CAL
LI INVALID CAL
These messages will be displayed when a specific electrode does not calibrate properly during a
calibration sequence. If more than six standard draws are required or the difference between STD
A and STD B voltages is out of range, the INVALID CALIBRATION message will be displayed
for each channel which exhibits either of these conditions. A copy of the calibration report
should be printed to provide the electrode output voltages measured during calibration. The
electrode voltage range table identifies correct voltage ranges for each electrode configuration. If
the difference between STD A and STD C is out of range for Li
+
or Na
+
, LI INVALID CAL will
be displayed.
Possible causes and remedies:
•
Check to ensure calibration reagents are transported correctly. Check for leaks, blockages,
bubbles, or improper pump flow rate.
•
Check or replace Fluid Pack
•
Check whether electrodes are plugged in properly. Check the slide mechanism to ensure
it is in the correct position and firmly pushed into electrode connectors. Clean and/or
replace electrode.
•
If A-C is out of range for Na+, repeat the calibration or replace the Fluid Pack
•
Check or replace CPU Card.
Summary of Contents for SmartLyte
Page 1: ......
Page 65: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 65 of 81 ...
Page 66: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 66 of 81 ...
Page 67: ...7 Illustrated Parts List ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 67 of 81 ...
Page 77: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 77 of 81 ...
Page 78: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 78 of 81 ...
Page 79: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 79 of 81 ...
Page 80: ...9 Schematics ECO 9085 SOP05 5038F Rev 05 Effective Date 12 06 18 Page 80 of 81 9 2 Printer ...