testing may be performed with the iRayA6s and the detector by IEC 61223-3-4, 5.8,
5.9.
Enable the iRayA6s and with the collimator perpendicular to the test detector, make
exposures into the test detector and capture the resulting data.
Compare the result with the factory release parameters (indicated in the cart below).
For results outside these parameters, discontinue use and contact DEXCOWIN.
Table 8-1. Test Acceptance Ranges
Test
Description
Acceptance
limits
Exposure Time (mSec)
50ms
100ms
300ms
1350ms
kVp
Accuracy
50kV ±10%
45kV
~ 55kV
45kV
~ 55kV
45kV
~ 55kV
45kV
~ 55kV
kVp
Accuracy
80kV ±10%
72kV
~ 88kV
72kV
~ 88kV
72kV
~ 88kV
72kV
~ 88kV
Timer
Accuracy
Set point
±10%
45mS
~ 55mS
90mS
~ 110mS
270mS
~ 330mS
1215mS
~ 1485mS