USER Manual
102
Table 11-34 Leakage radiation at 1[m] from the focal spot
DIRECTION
HORIZONTAL
VERTICAL
Section
Test 1
Test2
Test3
Test1
Test2
Test3
0˚(360˚)
ND
ND
ND
ND
ND
ND
45˚
ND
ND
ND
ND
ND
ND
90˚
ND
ND
ND
ND
ND
ND
165˚
ND
ND
ND
ND
ND
ND
180˚
ND
ND
ND
ND
ND
ND
225˚
ND
ND
ND
ND
ND
ND
270˚
ND
ND
ND
ND
ND
ND
315˚
ND
ND
ND
ND
ND
ND
Unit: [mR/h, (mGy/h)]
ND: Not Detected
11.7.2
Measurement in 1[cm] from case surface
The leakage radiation measured at 1 cm from surface of iRayA6s (ADX6000s) was
operated at 17 different points and the locations are as Figure 11.10 and the results
are as Table 11-35.
Unit: [mR/h, mGy/h]
The leakage radiation measured at 1 cm from surface is below 12 [mR/h] (0.12
[mGy/h]).
Summary of Contents for ADX6000s
Page 1: ......
Page 9: ...USER Manual 8 1 3 Symbols marked on iRayA6s...
Page 110: ...USER Manual 109...