6
b)
Technical data
Power supply............................................... :
100 / 110 / 115 / 120 / 230 / 240 VAC 50-60 Hz
Power consumption with rated load............ :
0.63 A AC - 3.15 A AC (depending on mains
voltage and load)
Permissible compressive force....................:
max. 7000 N (double-sided)
Permissible compressive force per side......:
max. 3500 N
Fuse............................................................ :
T 0.63 A – T 3.15 A (depending on mains voltage)
Operating mode at max. rated load.............:
intermittent duty AB 2 min./18 min. Protection
classification................................................ :
I or II
Noise level...................................................:
65 dB(A)
Current input of all slave drives .................:
max. 7.5 ADC
Drive type.................................................... :
dual drive
Drive options............................................... :
MONOMAT / DUOMAT / TRIOMAT /
TRIO/QUAD / QUADROMAT
Stroke
(other strokes on request)............. :
87 – 69 (standard) / 87 – 87 / 69 – 69 mm
Adjustment speed .....................................:
from 3 to 5.5 mm/s
Protection category......................................:
IP44
Colours........................................................ :
see sales brochure
Length x width of the drive.......................... :
750 x 175 mm
750 x 190 mm with "Emergency Release"
Axle base.....................................................:
581 mm
Weight......................................................... :
approx. 6.3 kg (without battery backup system)
Voltage........................................................ :
2 x 12 VDC
Capacity.......................................................:
1.2 Ah (with new batteries))
Fuse............................................................ :
T 15 A
Battery type................................................. :
Lead battery
Ambient and storage conditions
Room temperature...................................... :
from +10° to +40° C
Rel. humidity................................................:
from 30% to 75%
Barometric pressure.................................... :
from 700 hPa to 1060 hPa
≤
2
3)
4)
Dimensions and weights
Technical data of battery backup system
1)
2
3)
4)
Operating mode =
, i.e. run for 2 min. max. under rated load, then a rest
period of 18 min. must be observed, otherwise operational
!
According to regulations, no more than two drives may be run simultaneously at the rated load!
Stroke = maximum distance that the push block travels from the starting position in the retracted state to the end
position in the extended state.
Adjustment speed = speed at which the push block travels without load (speed varies according to the load)
intermittent duty AB 2 min./18 min.
failure could occur
)
DUOMAT 7 CARE L/HOSP L