16
Changing the mains connection lead and/or primary fuse
Disconnect the drive from the mains
Loosen the fastening screws
Carefully
Using a suitable tool,
flat plug
Change the mains connection cap with the mains connection lead or fuse(s)
Slot the flat plug
back on again.
with the flat
Retighten the fastening screws on the mains connection cap.
Work to mains connection lead and replacement of same may only be carried out by
specialist personnel holding the qualifications described on page 3 or by persons who have
taken part and successfully completed the corresponding training programs offered by
DEWERT. After correctly changing the mains connection lead, check for safety and proper
function.
!
!
!
!
!
!
!
!
(pull out the mains plug).
on the mains connection cap (see bottom of
page 16).
pull the mains connection cap out of housing (take care with flat gasket!).
pull the
of the mains connection lead from the
fuse cartridge.
of the mains connection lead
Slot the mains connection cap,
gasket, back on
again
correctly positioned
Fastening screws
DUOMAT 7 CARE L/HOSP L