8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
11
KEYPAD FUNCTIONS, CONT.
Basic Functions, Cont.
This is the
LOCK/RELEASE
key. It is used to lock and unlock the display. If
the HOLD feature was enabled during setup and calibration, pressing this key
(after obtaining a stable weight) will cause the MV1 to lock onto the weight.
Pressing this key again will unlock the display and return it to zero.
If the AUTO LOCK feature was enabled, the weight display will lock after
placing a load on the scale and obtaining a stable weight value. Pressing this
key will unlock the display and return the weight display to zero, or if another
load is applied to the scale, lock onto the new weight after obtaining a stable
weight value.
NOTE:
The lock feature is for non-commercial (NOT "Legal for
Trade") applications.
Body Mass Index (BMI) Functions
This is the
NET/GROSS
key. It is used to toggle between Net and Gross
weight modes. The selected mode is indicated by turning on the appropriate
annunciator on the display. Note that if no valid tare weight has been entered;
pressing this key will cause (
No Tare
) to be shown in the 14-segment display.
This is the
TARE
key. It is a dual function key. Pressing it, selecting a memory
location and pressing the
ENTER
key (Pushbutton Tare mode) will cause the
current gross weight to be stored as the new tare weight. Pressing this key,
entering a numeric weight value (Keypad Tare), selecting a memory location
and then pressing the
ENTER
key will cause the value entered to be stored as
the new tare weight. Either method will change the weight display to the net
weight display mode and turn on the
Net annunciator.
Pushbutton Tare Function:
Press the
TARE
key and then the
ID/HEIGHT
key, input a number between 0
and 99, and then press the
ENTER
key to store current Gross weight in a
memory location as a recallable tare weight.
Keypad Tare Function:
Press the
TARE
key, input the weight value, press the
ID/HEIGHT
key, input a
number between 0 and 99, and then press the
ENTER
key to store the
entered value in a memory location as a recallable tare weight.
This is the
ID/HEIGHT
key. This key is used to display and enter an
identification (ID) used in the ID Storage feature and to enter the height (feet
and inches or centimeters) for the Body Mass Index (BMI) calculation.
Pressing this key will change the display to show the currently selected
identification (ID) and allow the entry of a new ID. The identification (ID) can
be up to 14 alphanumeric characters.
To recall the stored tare weight from the memory location, press the
ID/HEIGHT
key, input a number between 1 and 99, and then press the
TARE
key.
NOTE:
Tare weights equal to or greater than scale capacity cannot be entered.
In addition, the keypad tare weight division value must be the same as the scale
division value. For example, a unit with .05 kg as the division value will display
WRONG GRAdS
if you attempt to enter 1.03 for the tare weight.