8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
27
CALIBRATION SETUP, CONT.
FFILTER
With
FILTER
displayed, press the
ENTER
key. The display will change to
FILTER MOdE
.
Proceed to the
FILTER MOdE
(Filter Type) parameter.
Otherwise, press the
ID/HEIGHT
key to advance to the
CALIbRATION
menu.
FILTER MOdE
(Filter Type)
The display will show
FILTER MOdE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0, 1, or 2.
0 = None Disabled, No Filtering
1 = IIR
Infinite Impulse Response – (slower response but better for higher
capacity scales)
2 = FIR
Finite Impulse Response – (faster response, better for filtering out
vibration)
NOTE:
The following prompt,
FILTER LEVEL
will only be displayed if you select option 1 or
2 for
FILTER MOdE
.
FILTER LEVEL
(Filter Level Selection –
Only if
FILTER MOdE
= 1
or
2
)
The display will show
FILTER LEVEL
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0, 1, or 2.
0 = Minimal Filtering
1 = Moderate Filtering
2 = Custom Filtering
NOTE:
The following prompts,
FILTER bREAk
(Filter Break Range) and
FILTER VALUE
(Filter Value) will only be displayed if you select 2 (Custom Filtering) for
FILTER LEVEL
.
FILTER bREAk
(Filter Break Range –
Only if FILTER LEVEL
= 2
)
`
The filter break range is a number from 1 to 64 that corresponds to the number of division
changes to break out of filtering.
The display will show
FILTER bREAk
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a
new value, and then press the
ENTER
key to save it.
Allowable values are 1 through 64.
FILTER VALUE
(Filter Value –
Only if FILTER LEVEL
= 2
)
The filter level is a number from 1 to 16 that corresponds to the level of filtering with 1 being
the least and 16 being the greatest filtering.
The display will show
FILTER VALUE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 1 through 16.