8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
25
CALIBRATION SETUP, CONT.
SSCALE
With
SCALE
displayed, press the
ENTER
key. The display will change to
CAPACITY
. Proceed
to the
CAPACITY
(Scale Capacity) parameter.
Otherwise, press the
ID/HEIGHT
key to advance to the
ZERO
menu.
CAPACITY
(Scale Capacity)
The display will show
CAPACITY
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 1 through 65,535.
OIML
(OIML European Requirements)
The display will show
OIML
and the current setting. If the value displayed is acceptable,
press the
ENTER
key to save it. Otherwise, use the numeric keys 0/NO or
1/YES
to select a
new setting, and then press the
ENTER
key to save it.
Allowable values are YES or no.
YES = OIML (European) specific
requirements are enabled
no = OIML (European) specific
requirements are disable
INTERVAL
(Interval Setting)
`
The display will show
INTERVAL
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 1, 2, or 5.
dECIMAL
(Decimal Point Position)
`
The display will show
dECIMAL
and the current setting. If the value displayed is acceptable,
press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new value, and
then press the
ENTER
key to save it.
Allowable values are 0, 1, 2, or 3.
0 = # # # # # #
1 = # # # # #
.
#
2 = # # # #
.
# #
3 = # # #
.
# # #
MOTION RANGE
(Motion Range)
`
The motion range is the number of divisions of change permitted before indicating unstable
(the STABLE annunciator turns off).
The display will show
MOTION RANGE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0 through 9.
SAMPLE RATE
(Sample Rate)
The display will show
SAMPLE RATE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
The sample rate may be set from a minimum of 1 sample per second to a maximum of 100
samples per second in one sample per second intervals.
Allowable values are 1 through 100.