MedVue MV1 Owner’s
54
APPENDIX B
(Connecting to a Welch Allyn CVSM Device)
This section describes the required settings of the MedVue Model MV1 Medical Weight
Analyzer in order to connect to a Welch Allyn CVSM device.
To enter Setup on the MedVue Model MV1 Medical Weight Analyzer, remove the two (2)
Philips head screws from the right end cap (as viewed from the back of the MV1) and remove
the end cap. Look for a small hole under the endcap with a red switch button behind it. It can
be found near the bottom edge of the Model MV1 Medical Weight Analyzer. Refer to the
CALIBRATION SETUP (Protected Setup Parameters) section for more information.
1. Press the switch with a paper clip or similar tool and you should see
SEtuP
on the top of
the display and
SCALE
at the bottom.
2. If it is desired for the MedVue to automatically send the weight, height, and BMI when
the weight is locked, the first setting that must be changed is the automatic lock feature
of the MedVue. While in setup, press the
ID/H
key until the display shows
LOCK
at the
bottom.
3. Press the
ENTER
key and you should see
HOLd MOdE
at the bottom of the MedVue
display. To enable the auto locking feature, change this setting to “2” and then press the
ENTER
key.
4. The next setting that appears is the
HOLd bANd
. Enter “10” for this setting and then
press the
ENTER
key. Press the
UNITS
key to return to the setup menu.
5. The next setting that must be correct in order for communication to work is the
USb
MOdE
. While in the setup menu, press the
ID/H
key until
USb PORT
is displayed at the
bottom of the MedVue display and then press the
ENTER
key.
6. Change the
USb MOdE
to “4” and then press the
ENTER
key.
You should now have all the correct settings in order to communicate with the Welch
Allyn device.
Auto-Lock Operation
If you have the MedVue set to auto-lock, simply apply weight to the scale. When the scale
settles and the weight locks, the weight, height, and BMI information will be transmitted.
The height and BMI do not lock with the weight. If the height was not entered in prior to
applying weight on the scale, only the weight will be transmitted to the Welch Allyn device.
If the operator enters in the height, either by using the digital height rod or entering the height
using the keypad, the height and calculated BMI information will be transmitted at that time.
Manual Operation
If you do not have the auto-lock enabled, pressing the
key on the MedVue keypad will
transmit the weight, height and BMI information.
Pressing the
key will clear the current height and BMI. If it becomes necessary to
retransmit the information, it will be necessary to enter the height again.