8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
36
CALIBRATION SETUP, CONT.
HHI RESOLUTION
(High-Resolution Mode)
The High-Resolution Mode feature will display the weight at 10 times the standard
resolution. For example, a MV1 configured for 400 lb x 0.2 lb will display weight in high
resolution mode at 400 lb x 0.02 lb.
NOTE:
If
dECIMAL=3
, the High-Resolution Mode is not
available.
The display will show
HI RESOLUTION
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise, use the numeric keys 0/NO or
1/YES
to select a new setting.
Allowable values are YES or no.
YES = If High-Resolution Weight
mode is desired
no = High-Resolution Weight
mode is not desired
If viewing the High-Resolution Weight mode
immediately
is desired:
1. With
HI RESOLUTION
displayed, press
1/YES
key.
2. Press the
ENTER
key.
3. The display will change to
RS232 MOdE
.
4. Press the
key twice.
5. The MV1 will reset and display the weight in High Resolution mode.
Otherwise, after selecting the value for the High-Resolution mode parameter, press the
ENTER
key to save the selection and advance to the
SERIAL PORT
,
RS232 MOdE
(RS232
Port Output Mode) parameter to continue Calibration Setup.