8555-M512-O1 Rev E
y
MedVue MV1 Owner’s
33
CALIBRATION SETUP, CONT.
(Non- Protected Setup Parameters)
SSYSTEM
With
SYSTEM
displayed, press the
ENTER
key. The display will change to
LANGUAGE
. Proceed
to the
LANGUAGE
(Language) parameter.
Otherwise, press the
ID/HEIGHT
key to advance to the
SERIAL PORT
menu.
LANGUAGE
(Language)
`
The display will show
LANGUAGE
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0, 1, 2, or 3.
0 = English
1= Spanish
2 = French
3 = German
NOTE: The default language is English. Changing the language at this
prompt will change all subsequent prompts and displays to the language
selected.
bACkLIGHT
(Backlight Intensity)
`
The display will show
bACkLIGHT
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise use the numeric keys to enter a new
value, and then press the
ENTER
key to save it.
Allowable values are 0 through 255.
0 = Backlight OFF
through
255 = Backlight at full intensity
NOTE: The backlight intensity setting will affect battery life. A lower intensity
setting will conserve power and maximize battery operation.
kEYPAd bEEP
(Allow Keypad Beep)
This setting determines whether the MV1 beeps each time a key is pressed.
The display will show
kEYPAd bEEP
and the current setting. If the value displayed is
acceptable, press the
ENTER
key to save it. Otherwise, use the numeric keys 0/NO or
1/YES
to select a new setting, and then press the
ENTER
key to save it.
Allowable values are YES or no.
YES = The MV1 beeps each time a
key is pressed.
no = MV1 does not beep when a key
is pressed.