background image

37

36

Note the meaning of below terms used in the manual: 
THERAPY = HEALTH IMPROVEMENT and DIAGNOSTICS = PROGNOSIS OF HEALTH STATUS

Note the meaning of below terms used in the manual: 

THERAPY = HEALTH IMPROVEMENT and DIAGNOSTICS = PROGNOSIS OF HEALTH STATUS

Setting filters in BRT

Generally, it is best not to use the filters during BRT effect. The function of the filters is to cut 

out LOW or HIGH signals – this may be required by the experienced specialist who is absolutely 
certain about in the degree of disability the person. 

It is best to not use the filters during BRT effect, unless one is absolutely certain about in the 

degree of disability the person.

There are basically two filters found on the BRT panel (Fig. 11) – one is the LOW filter which runs 

from 100Hz-1000Hz and the other is the HIGH filter which runs from 1kHz-10kHz. (1kHz=1000Hz)

If one presses the button above the LOW filter knob, and sets the filter knob to 100Hz, then one 

effectively cuts out the LOW frequencies up to 100 Hz and allowing the 101 through 1,000 Hz to get 

through – this one would want to do if the disease process being treated was ACUTE.

If one presses the button above the HIGH filter knob, then one cuts out the HIGH frequencies – if 

you set the knob to say 2 KHz, then you would be cutting out the first 2kHz and allowing all the 

other frequencies through – this one would want to do if the disease process being treated was 

CHRONIC.

In order to determine what filter setting to specify, we can determine this from the potency that 

we measure of the specific organ or microbe when using VRT testing. If we examine Diagram 17 

below this is made clearer.

Looking at Diagram 17, if we determine that one of the organs we are testing, say prostate is 

at the D3 potency, then we would adjust the frequency to 100 Hz in order to cut out the fir

st 100 Hz 

range as the potency D3 is higher than 100 Hz. This will allow frequencies between 101 and 1,000 

Hz to penetrate, but cut out the frequencies below 100 Hz.

Diagram 17: Choosing Filters Based on Potencies

1

2

3

4

5 6 7 8 9 10

2

2

3

4

5 6 7 8 9 10

3

2

3

4

5 6 7 8 9 10

4

2

3

4

5 6 7 8 9 10

5

Checking that BRT is working 

In order to verify if the bioresonance effect module is working OK, the Wellness module control 

panel is needed.

1. Switch on BRT mode with the «POWER» button on the «BRT» panel (see Fig. 11). In this case, 

the corresponding LED illuminates.

2. Set «VOLUME» to 60 units.
3. Switch on the «Control» mode with the «CONTROL» button. In this case, the corresponding 

LED illuminates and an audible signal sounds.

4. Switch on the «Inversion» mode with the «INVERSION» button. The sound will stop, signaling 

pathological information suppression mode. Switch on the «inversion» mode with the «INVERSION» 

button again.

5. Turn on the LF filter by pressing the «LOW» button of the «FILTERS» block. Adjustment of the 

«LOW» knob verifies the functioning of LF (the sound should fade when the knob is rotated in the 

direction of 100). Turn off the LF filter.

6. Turn on the HF filter by pressing the «HIGH» button of the «FILTERS» block. Adjustment of the 

«HIGH» knob verifies the functioning of HF (the sound should fade when the knob is rotated in the 

direction of 10). Turn off the HF filter.

7. Switch on the «Interrupt» mode with the «INTERRUPT» button. Set the «TIME OF INTERRUPT» 

knob to position «1». There should be a long beep and a short pause. At position «15» there should 

be a short beep and a long pause. Switch off «Interrupt» mode. 

8. Switch off «Control» mode.
9. Turn off BRT

Fig. 11: «BRT» panel

This is the end of the testing routine – this does not necessarily have to be repeated every time 

the specialist uses BRT – perhaps it is good to run this checking routine every day when BRT is 
began with visitors.

CHOOSING FILTERS BASED ON POTENCIES

Summary of Contents for DeVita Professional

Page 1: ...Note the meaning of below terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS OF HEALTH STATUS...

Page 2: ...Note the meaning of below terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS OF HEALTH STATUS...

Page 3: ...IX Manufacturer EC Declaration of Conformity CONTENT 1 INTRODUCTION 2 SPECIFICATION The measure mode Electroacupuncture effect mode Electromagnetic effect mode Electronic selector ES Quantum effect re...

Page 4: ...the measure mode measurements can be made according to the methods of Dr Voll and Dr Schimmel Parameters used are Conductivity indicator ranges from 0 to 100 Accuracy of measurement 1 Voltage between...

Page 5: ...SETUP The device can be set up relatively easily by following these simple instructions 1 Connect the AC adapter to the 12V socket on the device see Fig 3 number 2 The polarity of the voltage does no...

Page 6: ...e will refer to it as the Probe from here on first you must press the PROBE button on the MEASURE panel see Fig 4 number 4 In this case the PROB LED illuminates green Fig 4 number 2 Then on the THERAP...

Page 7: ...ECTING THE ELECTRONIC SELECTOR 2 Connect SEL to PASSIVE together with short wire 3 Connect PROBE to connector The procedure is summarized in Diagram 2 below 1 On the THERAPY panel set the THERAPY DIAG...

Page 8: ...resonator honeycomb with ampoules 1 On the measures panel press the PROBE button Fig 4 number 4 In this case the PROB LED illuminates 2 On the THERAPY panel set the THERAPY DIAGNOSTICS button to the D...

Page 9: ...y from the list then you simply enter the number zero 0 and it will be erased ELECTRONIC SELECTOR 6 2 ELECTRONIC SELECTOR ES On the PREPARATIONS display the message 01 N_ _ _ _ _ N should appear On th...

Page 10: ...did Voll and Schimmel We use a pressure of about 100 150 grams The pressure will differ according to age the condition of the skin whether they have psoriasis etc There are 10 fingers and we are free...

Page 11: ...rce for an electrode of 3 mm diameter 60 110 gram force for an electrode of 4 mm diameter 70 200 gram force The electrode we use on the Professional is usually 3 mm Moving to the measurement plateau t...

Page 12: ...and it is impossible to find any energy points 5 Could have very dry skin or skin affected with eczema or psoriasis 6 5 IS IT POSSIBLE TO TEST DIFFERENT THINGS AT THE SAME TIME In order not to kill t...

Page 13: ...left and right It is possible to identify whether organ systems have inflammation or degeneration using this method and then using Bioresonance VRT testing to identify the specific organs involved Se...

Page 14: ...the 4 measures or quadrants to identify pathology This will become clearer as we study the numbers on Diagram 10 below Diagram 9 Quadrant Measuring Relationships Specifically the top diagram shows an...

Page 15: ...ond electrode the harmonious frequencies being positively fed back positive feedback whilst the disharmonious frequencies are negatively fed back i e inverted negative feedback AN EXAMPLE OF INFLAMMAT...

Page 16: ...ng their healing process as well as inverting and neutralizing any pathological frequencies from the visitor and their body fluids Basic effect is carried out using the hand and foot electrodes before...

Page 17: ...T BRT into the left hand and the other hand electrode from the OUTPUT BRT into the right hand Energy will then flow from the strong to the weak therefore strengthening the right side of the body in th...

Page 18: ...ILS 1 Connect the honeycomb or resonator and hand electrode with the INPUT BRT connector body fluids 2 Connect resonator and hand electrode to OUTPUT BRT connector water alcohol 1 2 INVERSION ON INTER...

Page 19: ...ould adjust the frequency to 100 Hz in order to cut out the first 100 Hz range as the potency D3 is higher than 100 Hz This will allow frequencies between 101 and 1 000 Hz to penetrate but cut out the...

Page 20: ...use the ve polarity or button DOWN 8 When the button is DOWN on the Sinusidal wave we use this setting when there is no degenerative process but the problem is more functional See Diagram 18 for a su...

Page 21: ...s own frequencies Our most important energy source is light and the entire spectrum of colours is derived from light Sunlight which contains all the wavelengths consists of the entire electromagnetic...

Page 22: ...the skin at a speed of 1 2 cm sec In COMPLEX EFFECT MODE Quantum light is used half the time of Electromagnetic Effect 19 Zones can be used on body 7 5 COMPLEX EFFECT MODE COMPLEX EFFECT is a synthesi...

Page 23: ...t to transfer contaminated energetic information to the sugar pills Small plastic bag and sticker to place the homeopathics and label them for the visitor The potency of the homeopathic copy will be e...

Page 24: ...st 3 days in the visitor as this will get stored in the PLASMA MEMORY 4 Choose the remedies from the ELECTRONIC SELECTOR that you want on the various levels each remedy that you choose will have separ...

Page 25: ...accordance with the obligatory state standards effective technical documentation and classified as fit for operation Technical control check mark Seal here _____________________ ______________________...

Page 26: ...w terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS OF HEALTH STATUS Note the meaning of below terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS...

Page 27: ......

Reviews: