background image

19

18

Note the meaning of below terms used in the manual: 
THERAPY = HEALTH IMPROVEMENT and DIAGNOSTICS = PROGNOSIS OF HEALTH STATUS

Note the meaning of below terms used in the manual: 

THERAPY = HEALTH IMPROVEMENT and DIAGNOSTICS = PROGNOSIS OF HEALTH STATUS

6.3 SETTING UP FOR BIORESONANCE MEASURES

Before you begin bioresonance testing, you need to set up the Professional device as follows:

Connect a short length of wire on the side of the device from the SEL connector to the PASSIVE 

connector – you can use any of the two passive connectors, it does not matter as they are the 
same (see Fig. 2).

Connect the hand electrode to a long wire from the remaining PASSIVE connector.

Press the PROBE button (see Fig. 4).

On the SELECTOR section of the device, press POWER and LIGHT 

On the «THERAPY» panel of the device, where we see the THERAPY/DIAGNOSTICS switch, place 

this on the DIAGNOSTICS section.

You are now ready for bioresonance testing.

Search Techniques and Electrical Parameter Measurement of Biological

Active Points (BAP)

In order to be successful in bioresonance testing using the Probe, it is important to identify the 

correct point on the fingers. When one chooses an energetic point on the finger, in the middle of the 
finger just below the nail bed (see Fig. 9), it does not matter how hard one press with the probe, the 

conductivity indicator should show the same. This is the energetic point that interests us.

On a technical level, there are two points to bear in mind:
• We need to get the indicator correct by applying the correct pressure - this is known as 

«opening» the point or energizing the point.

• Then we can ask the question and get the answer through the indicator.

It is known as bioresonance as we use a voltage input into 

the energetic point (Voltage between the measure probe and 
hand electrode is 1.25 V, current is not greater than 12.5 mkA).

If the energetic signature is the same as there is no 

resonance, then there will not be an indicator drop. However, if 

there is resonance then there will be a stress on the Autonomic 

Nervous System and we will witness an indicator drop.

It is similar to the tuning fork phenomenon.

We use the mid-point of the fingers – not the acupoints as did Voll and Schimmel. We use a 

pressure of about 100 – 150 grams.

The pressure will differ according to age, the condition of the skin, whether they have psoriasis, 

etc. There are 10 fingers and we are free to choose whichever finger gives the best and most 

reliable reading.

We need to use the lightest pressure that gives us the best reading. Using too much pressure 

will result in «killing» the point and then we would need to calibrate with another point.

Always wet the probe point with a moistened cotton wool swab or use the swab directly on the 

finger to moisten it. Try to keep the point or probe slightly moist throughout testing. It is best to use 

a 0.1% saline solution to moisten the cotton wool ball.

If the indicator on the Voll setting (not VRT setting) gives us a reading above 20 unit then this 

is acceptable. If is lower than 15 unit, then we need to find another finger. We usually use the first 
three fingers, near the nail bed, in the middle of the finger. In these fingers and points there is 

minimal movement of the skin so will not interfere with our readings.

At the point where you place the measure probe, there will be a little hollow like a little «hole». 

This takes a little practice to master but persistence and patience is what is required to master this 
crucially important technique.

Press the finger with the probe three times to energize it – on all three times it must show 20 

unit or the same reading in order for it to be a reliable point. If there are differences between these 
three readings then this is not a reliable point and either you have to find another finger, or check 

the pressure of your measure probe – it should not be more than 100 – 150 g.

On the second time we can release the pressure to make certain that there is an indicator 

drop. The third time we can again increase the pressure and we should find that the indicator drop 
returns to the 20 unit level. If it does not then we would not use this point and find another finger.

The more one practices, the better the measurement will become as our muscles get trained 

to use the same pressure consistently.

When you have found the point on the Voll testing mode and the reading is above 20 unit, then 

you can press the second button on the measure probe while holding a firm pressure on the point 

to enter the VRT mode, the indicator should automatically go to 80 (see Fig. 5, button 2).

Fig. 9

Summary of Contents for DeVita Professional

Page 1: ...Note the meaning of below terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS OF HEALTH STATUS...

Page 2: ...Note the meaning of below terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS OF HEALTH STATUS...

Page 3: ...IX Manufacturer EC Declaration of Conformity CONTENT 1 INTRODUCTION 2 SPECIFICATION The measure mode Electroacupuncture effect mode Electromagnetic effect mode Electronic selector ES Quantum effect re...

Page 4: ...the measure mode measurements can be made according to the methods of Dr Voll and Dr Schimmel Parameters used are Conductivity indicator ranges from 0 to 100 Accuracy of measurement 1 Voltage between...

Page 5: ...SETUP The device can be set up relatively easily by following these simple instructions 1 Connect the AC adapter to the 12V socket on the device see Fig 3 number 2 The polarity of the voltage does no...

Page 6: ...e will refer to it as the Probe from here on first you must press the PROBE button on the MEASURE panel see Fig 4 number 4 In this case the PROB LED illuminates green Fig 4 number 2 Then on the THERAP...

Page 7: ...ECTING THE ELECTRONIC SELECTOR 2 Connect SEL to PASSIVE together with short wire 3 Connect PROBE to connector The procedure is summarized in Diagram 2 below 1 On the THERAPY panel set the THERAPY DIAG...

Page 8: ...resonator honeycomb with ampoules 1 On the measures panel press the PROBE button Fig 4 number 4 In this case the PROB LED illuminates 2 On the THERAPY panel set the THERAPY DIAGNOSTICS button to the D...

Page 9: ...y from the list then you simply enter the number zero 0 and it will be erased ELECTRONIC SELECTOR 6 2 ELECTRONIC SELECTOR ES On the PREPARATIONS display the message 01 N_ _ _ _ _ N should appear On th...

Page 10: ...did Voll and Schimmel We use a pressure of about 100 150 grams The pressure will differ according to age the condition of the skin whether they have psoriasis etc There are 10 fingers and we are free...

Page 11: ...rce for an electrode of 3 mm diameter 60 110 gram force for an electrode of 4 mm diameter 70 200 gram force The electrode we use on the Professional is usually 3 mm Moving to the measurement plateau t...

Page 12: ...and it is impossible to find any energy points 5 Could have very dry skin or skin affected with eczema or psoriasis 6 5 IS IT POSSIBLE TO TEST DIFFERENT THINGS AT THE SAME TIME In order not to kill t...

Page 13: ...left and right It is possible to identify whether organ systems have inflammation or degeneration using this method and then using Bioresonance VRT testing to identify the specific organs involved Se...

Page 14: ...the 4 measures or quadrants to identify pathology This will become clearer as we study the numbers on Diagram 10 below Diagram 9 Quadrant Measuring Relationships Specifically the top diagram shows an...

Page 15: ...ond electrode the harmonious frequencies being positively fed back positive feedback whilst the disharmonious frequencies are negatively fed back i e inverted negative feedback AN EXAMPLE OF INFLAMMAT...

Page 16: ...ng their healing process as well as inverting and neutralizing any pathological frequencies from the visitor and their body fluids Basic effect is carried out using the hand and foot electrodes before...

Page 17: ...T BRT into the left hand and the other hand electrode from the OUTPUT BRT into the right hand Energy will then flow from the strong to the weak therefore strengthening the right side of the body in th...

Page 18: ...ILS 1 Connect the honeycomb or resonator and hand electrode with the INPUT BRT connector body fluids 2 Connect resonator and hand electrode to OUTPUT BRT connector water alcohol 1 2 INVERSION ON INTER...

Page 19: ...ould adjust the frequency to 100 Hz in order to cut out the first 100 Hz range as the potency D3 is higher than 100 Hz This will allow frequencies between 101 and 1 000 Hz to penetrate but cut out the...

Page 20: ...use the ve polarity or button DOWN 8 When the button is DOWN on the Sinusidal wave we use this setting when there is no degenerative process but the problem is more functional See Diagram 18 for a su...

Page 21: ...s own frequencies Our most important energy source is light and the entire spectrum of colours is derived from light Sunlight which contains all the wavelengths consists of the entire electromagnetic...

Page 22: ...the skin at a speed of 1 2 cm sec In COMPLEX EFFECT MODE Quantum light is used half the time of Electromagnetic Effect 19 Zones can be used on body 7 5 COMPLEX EFFECT MODE COMPLEX EFFECT is a synthesi...

Page 23: ...t to transfer contaminated energetic information to the sugar pills Small plastic bag and sticker to place the homeopathics and label them for the visitor The potency of the homeopathic copy will be e...

Page 24: ...st 3 days in the visitor as this will get stored in the PLASMA MEMORY 4 Choose the remedies from the ELECTRONIC SELECTOR that you want on the various levels each remedy that you choose will have separ...

Page 25: ...accordance with the obligatory state standards effective technical documentation and classified as fit for operation Technical control check mark Seal here _____________________ ______________________...

Page 26: ...w terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS OF HEALTH STATUS Note the meaning of below terms used in the manual THERAPY HEALTH IMPROVEMENT and DIAGNOSTICS PROGNOSIS...

Page 27: ......

Reviews: