Cavitron™ PROPHY-JET® Air Polishing
Prophylaxis System: 100/115 VAC Systems
Mode of Operation: Continuous operation with intermittent loading:
10 minutes ON / 20 minutes OFF
Electrical Voltage 100V Gen 122J
115V Gen 122
Current
1.1/1.1 amps @ 100v/115v
Frequency
50/60 Hertz
Water Pressure 25–60 psig (172–414 kPa)
Air Pressure
65–100 psig (448–690 kPa)
Flow Rate
Minimum setting (CCW) 0–15 ml/min
Maximum setting (CW) >55 ml/min
Weight
12 lb. (5,4 kg)
Dimensions
Height:
4 in.
Width:
7
3/8
in.
Depth:
9
1/2
in.
Handpiece
Cable:
81
in.
Foot
Control
Cable:
96
in.
Power
Cord:
84
in.
Water Supply Line:
96 in.
Air Supply Line:
98 in.
Cavitron™ PROPHY-JET
®
Air Polishing
Prophylaxis System: 230 VAC Systems
Mode of Operation: Continuous operation with intermittent loading:
10 minutes ON / 20 minutes OFF
Electrical Voltage 230V Gen 122U
Current
.7/.6 amps @ 230v
Frequency
50/60 Hertz
Water Pressure 172–414 kPa
Air Pressure
448–690 kPa
Flow Rate
Minimum setting (CCW) 0–15 ml/min
Maximum setting (CW) >55 ml/min
Weight
5,4 kg
Dimensions
Height:
10,2 cm
Width:
18,7
cm
Depth:
24,1
cm
Handpiece
Cable:
2,0
m
Foot
Control
Cable:
2,4
m
Power
Cord:
2,1
m
Water Supply Line:
2,4 m
Air Supply Line:
2,5 m
OPERATING ENVIRONMENT
Temperature
10 to 25 degrees celsius
Relative Humidity
30 to 75 percent ( non-condensing )
TRANSPORT AND STORAGE CONDITIONS
Temperature
0 to 70 degrees celsius
Relative Humidity
10 to 95 percent ( non-condensing )
Atmospheric Pressure
500 HPa to 1060 Hpa
AC POWER
IPX1 classification of increase of water
TYPE B EQUIPMENT
FOOTSWITCH
Footswitch not for operating theatres
Protection Class- IPX1
Refer to instruction manual
for information
POWER ON
POWER OFF
Section 11: Specifications
Section 12: Classification
• Type of protection against electric shock:
Class 1
• Degree of protection against electric shock:
Type B
• Degree of protection against the harmfulingress of water:
Ordinary
• Mode of operation:
Continuous
• Degree of safety of application in the presence of a flammable
anaesthetic mixture with air or with oxygen or nitrous oxide:
• According to medical device directive:
IIA (rule 9)
Section 13: Disposal of Unit
• Accordance with local and state laws.
Equipment not suitable for use in the presence
of flammable anesthetics or oxygen.