Appendix A: Electromagnetic Emissions and Immunity
o) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal
to 16 A / phase and ME EQUIPMENT and ME SYSTEMS with RATED input current greater than
16 A / phase.
p) Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less than or equal
to 16 A / phase.
q) At some phase angles, applying this test to ME EQUIPMENT with transformer mains power
input might cause an overcurrent protection device to open. This can occur due to magnetic flux
saturation of the transformer core after the voltage dip. If this occurs, the ME EQUIPMENT or ME
SYSTEM shall provide BASIC SAFETY during and after the test.
r) For ME EQUIPMENT and ME SYSTEMS that have multiple voltage settings or auto ranging
voltage capability, the test shall be performed at the minimum and maximum RATED input voltage.
ME EQUIPMENT and ME SYSTEMS with a RATED input voltage range of less than 25 % of the
highest RATED input voltage shall be tested at one RATED input voltage within the range. See
Table 1 Note c) for examples calculations.