Appendix A: Electromagnetic Emissions and Immunity
Compliance with IEC 60601-1-2:2014
Propex IQ
®
Apex Locator does not have Essential Performance.
Use of
Propex IQ
®
Apex Locator adjacent to, or stacked with other equipment should
be avoided because it could result in improper operation. If such use is necessary,
Propex IQ
®
Apex Locator and the other equipment should be observed to verify that
they are operating normally.
WARNING: Portable RF communication equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) from any part of
Propex IQ
®
Apex Locator. Otherwise, degradation of the
performance of this equipment could result.
Guidance and manufacturer’s declaration - electromagnetic emissions
Propex IQ
®
Apex Locator is intended for use in the electromagnetic environment specified below.
The customer or the user of
Propex IQ
®
Apex Locator should assure that it is used in such an
environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF
Emissions
CISPR 11
Group 1
Propex IQ
®
Apex Locator generates and uses
conductively coupled radio-frequency energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF
Emissions
CISPR 11
Class B
Propex IQ
®
Apex Locator is suitable for use in
establishments directly connected to low voltage
power supply network which supplies buildings used
for domestic purposes.
Professional healthcare facility environment.
Harmonic Emissions
IEC 61000-3-2
Not applicable
-
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable
-