![Dentsply Sirona Propex IQ Directions For Use Manual Download Page 68](http://html1.mh-extra.com/html/dentsply-sirona/propex-iq/propex-iq_directions-for-use-manual_2483689068.webp)
Appendix A: Electromagnetic Emissions and Immunity
68/74
BENPPIQDFUWEB / Rev.04 / 08-2018
Recommended separation distances between portable and mobile RF communications equipment
and the
Propex IQ
®
Apex Locator
The
Propex IQ
®
Apex Locator is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
Propex IQ
®
Apex
Locator can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
Propex IQ
®
Apex Locator as recommended below, according to the maximum output power of the
communication equipment.
Rated
maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12.0
12.0
23.0
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance to
decrease the likelihood that mobile / portable communications equipment could cause interference
if it is inadvertently brought into patient areas.
d
1.2
P
×
=
d
1.2
P
×
=
d
2.3
P
×
=