Technical Characteristics
Material
Single use
Propex IQ
®
Disposable
Barrier Sleeve: Polyethylene
Propex IQ
®
Clamp:
Stainless steel
Rubber pad:
SEBS
Threaded insert: brass
Grub screw: stainless steel A1
Power supply
Rechargeable 2x AAA Ni-MH
Batteries:
Minimum capacity 750 mAh
1.2 VDC nominal;
INTERNALLY POWERED DEVICE
Charger (type: FW7662M/05)
•
Power supply:
100 - 240 VAC
•
Frequency:
50 - 60 Hz
•
Nominal power output:
5.5 VA
•
Electrical safety class:
Class II
Applied part
BF (
Propex IQ
®
File Clip and
Propex IQ
®
Lip Clip)
Protection against water ingress
IP 20
Safety level in the presence of flammable anesthetic
gas mixtures or oxygen
Not suitable for use in the presence of
flammable anesthetic gas mixtures or
oxygen
Ambient conditions
•
Use: in enclosed spaces
•
Ambient temperature:
15°C - 35°C
(59°F - 95°F)
•
Relative humidity:< 80%;
non-condensing at 0°C (32°F)
•
Operating altitude:
< 2 000 m (approximately 6561 ft)
above sea level
•
Maximum housing temperature
according to IEC 60601-1/A1:2012
Specification