![Dentsply Sirona Propex IQ Directions For Use Manual Download Page 71](http://html1.mh-extra.com/html/dentsply-sirona/propex-iq/propex-iq_directions-for-use-manual_2483689071.webp)
BENPPIQDFUWEB / Rev.04 / 08-2018
71/74
Appendix A: Electromagnetic Emissions and Immunity
Input a.c. power port
PHENOMENON
BASIC EMC
STANDARD OR TEST
METHOD
IMMUNITY TEST LEVELS
PROFESSIONAL
HEALTHCARE FACILITY
ENVIRONMENT
TEST RESULTS
Electrical fast
transients / bursts
a) l) o)
IEC 61000-4-4
± 2 kV
100 kHz repetition
frequency
Pass
Surges
a) b) j) o)
Line-to-line
IEC 61000-4-5
± 0,5 kV, ±1 kV
Pass
Surges
a) b) j) k) o)
Line-to-ground
IEC 61000-4-5
± 0,5 kV, ± 1 kV, ± 2 kV
Pass
Conducted
disturbances induced
by RF fields
c) d) o)
IEC 61000-4-6
3 V
m)
0,15 MHz - 80 MHz
6 V
m)
in ISM bands
between 0,15 MHz and
80 MHz
n)
80% AM at 1 kHz
e)
Pass
Voltage dips
f) p) r)
IEC 61000-4-11
0% UT; 0,5 cycle
g)
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
q)
Pass
0% UT; 1 cycle and
70% UT; 25/30 cycles
h)
Single phase: at 0°
Pass
Voltage interruptions
f) i) o) r)
IEC 61000-4-11
Pass