9
Dentsply Sirona
1 Important information
Service Manual inFire HTC speed
1.3 Safety information
båÖäáëÜ
Wireless phone interference with
laboratory equipment
To ensure safe operation of laboratory equipment, the use of mobile
wireless phones in laboratory environments must be prohibited.
Note on the prevention, recognition and
elimination of unintended
electromagnetic effects
This device is a piece of electrical equipment with a supply voltage below
1000 VAC and is intended for commercial use. It is to be installed in
dental laboratories or other areas with a controlled electromagnetic
environment. The applicable EMC requirements comply with EN
61326:2006.
Troubleshooting:
If you encounter difficulties, check the Troubleshooting section first and
proceed as instructed there.
WARNING
Risk of burns due to hot surfaces
Once the program has ended and the furnace has shut down, the
cooling fan will run until the interior temperature of the furnace has been
reduced to normal. The power plug may not be disconnected prior to this
point.
WARNING
Risk of burns due to objects falling out
As there is a serious risk of burns in this case, it is essential to install a
fire-proof table and floor covering.
I. For safety reasons, the furnace must be operated when standing. Only
in this way is it possible for the user to quickly avoid parts that fall out.
II. When using super speed sintering, wearing heat-resistant protective
clothing is recommended. In accordance with professional association
regulations, the operator of the site is responsible for providing the user
with appropriate protective clothing.
III. The operator of the site is responsible for giving the user appropriate
safety instructions.
IV. If a fracture occurs during the sintering process as a result of a failure
to adhere to the recommended drying procedure, there is a risk that hot
objects will fall out as soon as the furnace door is opened.
V. The furnace should only be loaded and unloaded using the crucible
fork. For safety reasons, a trained user must be present when the
furnace is being loaded or unloaded.