9 Replacing parts
Service Manual inFire HTC speed
9.8 Replacing the sintering chamber
båÖäáëÜ
5. Carefully release all terminals (A) on the contacts of the heating
elements (C).
6. Pull the connections (B) between the heating elements (C) up.
7. The heating elements (C) can now be removed up and out of the
sintering chamber. Place the heating elements (C) on a clean surface
and make sure that they are not contaminated.
8. Loosen the screw (D) on the temperature sensor (E).
9. Carefully remove the temperature sensor (E) with the mounting
device out of the sintering chamber guide.
C
B
A
A
NOTICE
Pull the heating elements vertically upward making sure that they do not
tilt. Do not apply any bending load to the heating elements; otherwise
they will break.
D
E
NOTICE
Pull the temperature sensor vertically upward making sure that it does
not tilt. The temperature sensor is very fragile and can break if tilted.