3
I nstructions — English
•Prior to initial use - always read the instructions for use.
•Only for US: prescription use – for use by dental clinicians only.
Indication for use
•Adjusting initial, finishing and polishing of dental surfaces and restorations.
Contraindication
•No known contraindications.
Precaution/Warning
Handpiece
Cross-contamination
•Clean, disinfect and sterilize the handpiece before first use and between uses.
•Follow the instructions in the instructions for use. The instructions are validated by the manufacturer.
Thermal
•Do not exceed 10,000 rpm.
•Verify the recommended rpm on the motor before start using the handpiece.
•Lubricate the handpiece between each use, debris can cause clogging and excessive friction.
Small objects
•Before start working: Check for secure seating by pulling the tip with your fingers. A firm resistance should be sensed.
•If tip comes out easily, try another tip or send handpiece for service.
Malfunction
•Do not disassemble any part of the handpiece.
•Lubricate the handpiece between each use, debris can cause clogging and excessive friction.
Vibration
•Discomfort for patient and dentist if speed is lower than 5,000 rpm.
Working tips
Cross-contamination
•Clean, disinfect and sterilize between uses.
•Do not re-use the EVA-tips. They cannot be cleaned properly.
EN
Summary of Contents for Profin pdx
Page 193: ...193 Notes ...