HD2114P/HD2134P
- 21 -
V2.3
INSTRUMENT
SIGNALS
AND
FAULTS
The following table lists all error indications and information displayed by the instrument and
supplied to the user in different operating situations.
Display indications
Explanation
ERR
This appears if the pressure sensor detects a value exceeding the limit
of 125% of the bottom scale nominal value.
OVER
Measurement overflow: this appears if the pressure sensor exceeds the
limit of 120% of the bottom scale nominal value. Over 125%, the
display indicates ERR.
Measurement overflow: this appears if the external temperature probe
is measuring a value exceeding the set measuring range.
LOG
MEM
FULL
Memory full; the instrument cannot store further data, the memory
space is exhausted.
SYS
ERR
#
Instrument management program error. Contact the instrument's
supplier and communicate the numeric code # reported by the display.
CAL
LOST
Program error: it appears after turning on for a few seconds. Contact
the instrument's supplier.
BATT TOO LOW
CHNG NOW
Indication of insufficient battery charge appearing on turning on. The
instrument issues a long beep and turns off. Replace the batteries.