HD2114P/HD2134P
- 13 -
V2.3
SERIAL/EraseLOG key - only HD2114P.2 and HD2134P.2
In measurement mode, this function starts and stops the data transfer to the serial output.
According to the settings entered in the
Print and log interval
menu item, a single sample can be
printed if
Print and log interval
=0 or a continuous indefinite printing of the measured data can be
set up if
Print and log interval
=1…3600.
The printing operation is accompanied by the display of the RS232 symbol and the blinking of the
battery symbol;
when using an external power supply the battery symbol does not appear
.
Press SERIAL to end the continuous printing.
Before starting the printing via the RS232C port, set the baud rate. To do so, select the
Baud Rate
menu item and select the maximum value equal to 38400 baud by using the arrows. Confirm by
pressing ENTER.
The DeltaLog9 software for PC will automatically set the baud rate value during connection.
If you
are using a different program than DeltaLog9, be sure the baud rate is the same for both the
instrument and the PC: the communication will only work in this way.
The USB 2.0 connection does not require the baud rate setting as it is automatically managed by the
application.
>>>
Erase memory - only HD2114P.2 and HD2134P.2
Press the MENU key until the ">>>_LOG_DUMP_or_ERAS" item is displayed and then press on
the SERIAL/EraseLOG key: this clears
permanently
all the data contained in the instrument's
memory.