HD2114P/HD2134P

 

-   11   -

 

V2.3

 

Use the arrows and confirm with ENTER. Please see the paragraph dedicated to wind speed 
measurement on page 14. 

4)

 

Pres Baro (barometric pressure): 

the wind speed detected by the Pitot tube is the result of 

different factors. Among these is the atmospheric pressure resulting from the formula 
reported on page 15. Using the arrows, enter the atmospheric pressure value present during 
the measurement. If the current pressure is not known, leave the parameter at the default 
value of 1013mbar. 

5)

 

Pres Stat (Static pressure)

: static pressure present in the measured duct. It is referred to the 

atmospheric pressure and is expressed in mbar. If the duct is open, in contact with the 
atmosphere, its value is set to zero (default value). If the duct is closed, the pressure 
difference compared to the atmospheric pressure, expressed in mbar, must be entered. 

In 

order to avoid the sensor breakage, do not use the Pitot tube if the static pressure 
exceeds the overpressure limit reported in the technical characteristics.

 

6)

 

Cost_PIT_DARC (constant of Pitot or Darcy tube)

:

 

using the arrows, enter the tube 

constant value. The value can be set from 0.80 to 1.20. If the constant is not known, enter 
1.00 for Pitot tubes and 0.84 for Darcy tubes. The set value is displayed when the instrument 
is turned on. 

7)

 

AVG TIME SECS

:  

time interval according to which the moving average is calculated

,

 

in seconds, during measurement of the flow. The value ranges from 1 (no average) to 99 
seconds. Please see the paragraph dedicated to wind speed measurement on page 14. 

8)

 

Management of memorized data (only HD2114P.2 and HD2134P.2)

: the message 

"

>>>_LOG_DUMP_or_ERAS

(Transfer data or erase)

 is scrolled in the comment line. 

The center figure reports the number of free memory pages (FREE). All memory data are 
permanently erased by pressing SERIAL/EraseLOG. By pressing LOG/DumpLOG, the data 
transfer of the logged data on the serial port is started: the "BAUD-RATE" must have 
previously been set to the maximum value (please see the menu items described below and 
the paragraph "STORING AND TRANSFERRING DATA TO A PERSONAL 
COMPUTER" on page 27).   

9)

 

Print and log interval (only HD2114P.2 and HD2134P.2)

: sets the interval in seconds 

between two loggings or data transfers to the serial port. The interval can be set at 0, 1s, 5s, 
10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s 
(20min), 1800s (30min) and 3600s (1hour). 

If the value 0 is set, SERIAL works on 

command: the sending of data to the serial port is performed each time the key is 
pressed

. Recording (LOG) is performed with one second intervals even if the interval is set 

to 0. With an interval from 1 to 3600s, continuous data transfer is started when the SERIAL 
key is pressed. To end the recording (LOG) and 

continuous

 data transfer operations 

(SERIAL with an interval greater than 0), press the same key again. 

10)

 

Sleep_Mode_LOG (Automatic turning off during recording) (only HD2114P.2 and 

HD2134P.2): 

this function controls the instrument's automatic turning off during logging, 

occurring between the capture of a sample and the next one. When the interval is lower than 
60 seconds, the instrument will always remain on. With intervals greater than or equal to 60 
seconds, it is possible to turn off the instrument between loggings: it will turn on at the 
moment of sampling and will turn off immediately afterwards, thus increasing the battery 
life. Using the arrows select 

YES

 and confirm using 

ENTER

 in order to enable the 

automatic turning off, select 

NO

 and confirm to disable it and keep the instrument on 

continuously.  

Note

: even if 

Sleep_Mode_LOG=YES 

is selected, the instrument does not turn off for less 

than one minute intervals. 

Summary of Contents for HD2114P.0

Page 1: ...Operating manual Manometers for Pitot tubes thermometers HD2114P 0 HD2114P 2 HD2134P 0 HD2134P 2 www deltaohm com Companies Brands of GHM English Keep for future reference...

Page 2: ...25 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 27 THE LOGGING FUNCTION ONLY FOR HD2114P 2 AND HD2134P 2 27 CLEARING THE MEMORY ONLY FOR HD2114P 2 AND HD2134P 2 27 THE PRINT FUNCTION ONLY FOR...

Page 3: ...uments are dataloggers They memorize up to 36 000 samples which can be transferred from the instrument connected to a PC via the RS232C serial port or USB 2 0 port The storing interval printing and ba...

Page 4: ...HD2114P HD2134P 4 V2 3 HD2114P 0 HD2134P 0 Micromanometer with Pitot tube Thermometer...

Page 5: ...it from the menu 10 C F ESC key when the thermocouple probe is not connected it allows manual modification of the temperature When double pressed changes the unit of measurement for temperature from d...

Page 6: ...HD2114P HD2134P 6 V2 3 HD2114P 2 HD2134P 2 Micromanometer with Pitot tube Thermometer...

Page 7: ...starts the data transfer from the instrument s memory to the PC 12 Mini USB type B connector for USB 2 0 For the connection to PC with cable CP23 13 MENU key allows access to and exit from the menu 14...

Page 8: ...ables all display segments for a few seconds displays the setting of the Pitot or Darcy tube constant please see the Cost_PIT_DARC parameter on page 11 and sets the instrument ready for normal measure...

Page 9: ...ferred to the PC HOLD key It increases the current parameter when used in the menu when used in measurement mode it freezes the measurement in progress and upon application of pressure on the key the...

Page 10: ...nt variable value MENU Key The first menu item is accessed by initially pressing on the MENU key press ENTER to go to the following items To modify the item displayed use the arrow keys and The curren...

Page 11: ...ber of free memory pages FREE All memory data are permanently erased by pressing SERIAL EraseLOG By pressing LOG DumpLOG the data transfer of the logged data on the serial port is started the BAUD RAT...

Page 12: ...r only works if the instrument and PC baud rates are the same If the USB connection is used this parameter value is automatically set please see the details on page 27 LOG DumpLOG key only HD2114P 2 a...

Page 13: ...tinuous printing Before starting the printing via the RS232C port set the baud rate To do so select the Baud Rate menu item and select the maximum value equal to 38400 baud by using the arrows Confirm...

Page 14: ...models measure air temperature through the thermocouple K The measurement provided by the instruments are Pv differential pressure wind speed calculated flow rate air temperature detected by the therm...

Page 15: ...ature detected by the thermocouple K present in some Pitot tube models If the tube is not fitted with one an external thermocouple K probe with miniature connector can be attached at the Pitot tube in...

Page 16: ...ent of less than 0 5 DIMENSIONS OF PITOT TUBES T1 T2 T3 T4 d L t Diameter d mm 3 5 8 10 Tip length t mm 33 55 88 135 Length L mm 300 400 600 500 800 500 800 1000 Order Code T1 300 T2 400 T2 600 T3 500...

Page 17: ...speed and therefore the calculated flow rate along the entire section and not on a single measurement point Procedure Using the UNIT key select the speed or flow rate unit of measurement for which you...

Page 18: ...ar 200mbar Maximum overpressure 300mbar 1bar Some units of measurement require a degree of multiplication the 3 symbol indicates the displayed value must be divided by 1 000 the 3 and 6 symbols indica...

Page 19: ...e tip probe In order to perform a correct contact measurement the measurement surface must be even and smooth and the probe must be perpendicular to the measurement plane A contact measurement is hard...

Page 20: ...er to improve contact and reduce reading time Whatever you do do not use water or solvent for this purpose A contact measurement is always very hard to perform It has high levels of uncertainty and de...

Page 21: ...ominal value Over 125 the display indicates ERR Measurement overflow this appears if the external temperature probe is measuring a value exceeding the set measuring range LOG MEM FULL Memory full the...

Page 22: ...ogress ERR error FUNC CLR max min and average values clearing FUNC CLRD max min and average values cleared HOUR hour LOG In PROG logging in progress LOG MEM FULL memory full LOG_CLRD memory data clear...

Page 23: ...eries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2114P 2 or HD2134P 2 are logging and battery voltage falls below the minimum operating l...

Page 24: ...ries leaking Always use good quality leakproof alkaline batteries Sometimes on the market it is possible to find new batteries with an insufficient charge capacity INSTRUMENT STORAGE Instrument storag...

Page 25: ...ating system Cable Instrument port PC port Installation of USB drivers HD2110CSNM RS232 MiniDin RS232 9 pole SubD No C 206 RS232 MiniDin USB Yes CP23 USB Mini USB USB No The instrument standard serial...

Page 26: ...nction KP Auto power off function ENABLE KQ Auto power off function DISABLE KZ Differential probes reset function RA Reading of LOG PRINT interval set RP 600 Battery level Resolut 0 01V RUA U Pa Main...

Page 27: ...e LOG ON The data stored in the memory can be transferred to a PC using the DUMP LOG command MENU key using ENTER select the _LOG_DUMP_or_ERAS item and then press on the LOG DumpLog key During data tr...

Page 28: ...orks with a standard serial communication program e g Hyperterminal only through the RS232 serial port of the instrument using the cable HD2110CSNM or the cable C 206 Connect the HD40 1 printer using...

Page 29: ...free RS232C COM or USB serial port of the PC 3 Turn on the instrument and set the baud rate to 38400 MENU ENTER until the Baud Rate parameter select 38400 using the arrows confirm with ENTER The param...

Page 30: ...operational safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store t...

Page 31: ...ac 12Vdc 1A Measuring units C F Pa mbar mmH2O PSI m s km h ft min mph knot l s m3 h cfm Security of memorized data Unlimited independent of battery charge conditions Time Date and time Schedule in rea...

Page 32: ...e Differential pressure 20mbar 200mbar Speed 2 55m s 3 180m s Temperature using thermocouple K 200 1370 C 200 1370 C Temperature using Pitot tube 200 600 C 200 600 C Maximum overpressure 300mbar 1bar...

Page 33: ...nces refer to the operating temperature expected for the thermocouple in agreement with the diameter of the thermoelements TOLERANCE CLASSES OF THERMOCOUPLES Tolerances according to IEC 60584 2 standa...

Page 34: ...n USB type A With integrated RS232 USB converter CP23 Connection cable Mini USB type B USB type A DeltaLog9 Software for transfer and management of the data on PC using Windows from 98 operating syste...

Page 35: ......

Page 36: ...s is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic equipment marke...

Page 37: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Page 38: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V2 3 28 08 2017...

Reviews: