In the temperature measurement by
penetration
the probe tip must be inserted to a depth of at least
60mm, the hot junction is housed in the end part of the probe. When measuring the temperature on
frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip
probe.
In order to perform a correct
contact
measurement, the measurement surface must be even and
smooth, and the probe must be perpendicular to the measurement plane. A contact measurement is
hard to perform due to various factors: the operator must be experienced in handling the probe and
consider all the factors influencing it.
So as to obtain the correct measurement, the insertion of a drop of oil or heat-conductive
paste is useful (do not use water or solvents). This method also improves the response time.