Rev 2022-03-08
Pharmafill Model TC4 User Manual
Deitz Company Inc.
TC4 User Manual
Page
34
9.3.
CLEAN-IN-PLACE CONTACT PARTS
The following must be cleaned in place unless further disassembly is done using tools. Contact
the factory if you need assistance.
1.
Hopper (optional, if equipped) – wipe down inside surfaces.
2.
Feeder pan (optional, if equipped) - wipe down inside surfaces.
3.
Inner chute – wipe down the all visible inside surfaces
4.
Divider Flag
5.
Counting head Back Plate
9.4.
NON-CONTACT AREAS
1.
Feeder pan dust chute and dust box – clean out dust box by removing four screws from
the rear cover and wiping or blowing with compressed air (wear eye protection, please).
2.
All other surfaces – follow guidelines in the next section.
9.5.
CLEANING RECOMMENDATIONS
Washdown
This machine is not waterproof and is not intended for full wash down. If full washdown is
performed on the equipment near the machine, it must be completely protected by a
waterproof cover or by other means. Washdown will void the warranty.
Cleaning solutions
Glass and stainless steel are resistant to most cleaning solutions. Other contact materials
such as aluminum and nonmetallics (plastics, or rubber) are generally less corrosion-resistant
and care should be exercised in their cleaning. Aluminum is readily attacked by acids as well
as highly alkaline cleaners, which can render the surface non-cleanable. Plastics are subject
to stress cracking and clouding from prolonged exposure to corrosive cleaning agents. Use a
USDA approved sanitizing solution that is safe for all materials listed below, in a spray
bottle, by lightly wiping down all contact surfaces. In the absence of such a cleaner,
recommendations follow.
Recommendations
Glass (tempered): This material is resistant to damage from most cleaners. Routine cleaning
can be done with soap and water, alcohol or acetone.