Rev 2022-03-08
Pharmafill Model TC4 User Manual
Deitz Company Inc.
TC4 User Manual
Page
15
7.3.3.
SETTING THE THRESHOLD AND ONE-SHOT DELAY ADJUSTMENTS (V2.0)
THRESHOLD (Factory setting = 8)
To adjust:
❑
Press and r (plus)
▪
Light bar will change to indicate current threshold
▪
“LO” and “DO” lights will flash alternately
❑
Press + to increase the threshold (less sensitive)
❑
Press – to reduce the threshold (more sensitive)
Will automatically exit adjustment mode after 4 seconds of inactivity
Will automatically reset light level for maximum performance before returning to normal.
ONE-SHOT DELAY (Factory setting = 1)
To adjust:
❑
Press and release – (minus)
▪
Light bar will change to indicate current one-shot delay
▪
“Clock” light will flash
❑
Press + to increase the one-shot delay.
❑
Press – to reduce the one-shot delay.
Will automatically exit adjustment mode after 4 seconds of inactivity
Will automatically reset light level for maximum performance before returning to normal.
+ (PLUS) BUTTON
- (MINUS) BUTTON
LIGHT BAR 1
LIGHT BAR 8
COUNT SIGNAL (YELLOW)
POWER ON (GREEN)
HS
(CLOCK)
DO
LO