Page
6
3. Technical data PM 20
External supply
- Operating voltage
- Operating current
Displayed only
during measurement
Battery
(not included in delivery)
Test parameters
- Test voltage
- Measuring range
- Resolution
- Test current monitoring
system
Test period
Test output sockets
Measured value indicator
Number of individual tests in
battery operation
Ambient temperature during
operation
Accessory included in delivery
Serial number
Test adapter
(not included in delivery)
8 ... 12 V DC
1 mA
20 ... 200 mA
9 V IEC6F22 / NEMA-1604-
"alcaline" type preferred
Max. 1200 V DC
- Rise speed
1000 V / sec.
Up to 1100 V
1 V
Automatic recognition whether:
-Limitation to reference value
Þ
Test current constant at 1 mA
-Response via ignition
Þ
Switch off the test voltage
Max. 1,5 sec
Until the measurement value is
displayed
Safety pole terminals (4 mm)
Negative pole: black
Positive pole: red
LCD, alpha/ numeric, 8-digit
Typically 2000
+10 to +35°C
2 test cables (1 m each)
2 safety test clips
1 power supply unit (230 V AC)
1 storage bag
Individual
(attached to the front side)
PA-DRL
Part No. 910 507
(for testing DPL 1/10... and DRL...)
PA-BXT
Part No. 910 508
(for testing BLITZDUCTOR XT and CT)