11.2 Battery Packs
Use only Defibtech battery packs in the DDU-2000 Series AED.
DBP-2003 and DBP-2013 Battery Packs
Category
Specification
Model number
DBP-2003
DBP-2013 (Aviation;TSO C-142a)
Main battery type
12VDC, 2800 mAh, Lithium/Manganese Dioxide.
Disposable, recyclable, non-rechargeable.
Capacity
125 shocks or 8 hours of continuous operation.*
Charge time
AED Mode:
• 4 seconds or less (from shock advised)*
Manual Mode:
• 9 seconds or less (150 J)**
• 12 seconds or less (200 J)**
Standby-life (installed in unit)
4 years*
*Typical, new battery, at 25°C
**typical, new battery depleted by 6 shocks, at 25°C
11.3 Self-Adhesive Defibrillation Pads
Use only Defibtech defibrillation pads with your DDU-2000 Series AED. Defibtech self-adhesive defibrillation pads
have the following characteristics:
Model number
DDP-2001
DDP-2002 and DDP-2003
Type
Adult
Child/Infant < 8 years
< 55 lbs. (25kgs)
Intended use
Disposable
Disposable
Adhesion
Self-adhesive
Self-adhesive
Active gel surface area
77 cm
2
each (nominal)
50 cm
2
each (nominal)
Cable/connector type
Integrated
Integrated
Cable length
122 cm (typical)
122 cm (typical)
Expiration date
2.5 years from date of manufacture
2.5 years from date of manufacture
Note
: In the event of a suspected pad defect, the pads should be clearly marked “Not for Use” and returned
to Defibtech, LLC for analysis. (Refer to
of this manual for contact information regarding returns.)
11.4 ECG Monitoring Adapter / Cable (Optional)
For use only with DDU-2400 and DDU-2450 AEDs.
Model number
DAC-2020 and DAC-2021
Patient connection
Type CF, fully defibrillation-protected
Cable length
2 meters
Patient lead wire designation
Lead type
DAC-2020
(AHA)
DAC-2021
(IEC)
Placement
Positive
Red - LL
Green - F
Left leg
Negative
White - RA
Red - R
Right arm
Reference
Black - LA
Yellow - L
Left arm
Typical lead connection
Lead II
Performance with
DDU-2400/2450 AED
for DDU-2000 Series.
11
.
Tec
hnical Specifications