Voice
Text
“SHOCK button not pressed”
Button Not Pressed
Purpose:
After shock is advised, the DDU-2000 Series AED will prompt user to press the flashing shock button.
If after 30 seconds the shock button is not pressed, the DDU-2000 Series AED will give this prompt and
immediately go to CPR mode.
No Shock Required Prompts
Voice
Text
“No shock advised”
“It is safe to touch the patient”
No Shock Advised
OK to Touch Patient
Purpose:
The DDU-2000 Series AED has determined that a shock is not required. The unit will not charge and
the SHOCK button will not be enabled. The user will be prompted to begin CPR.
CPR Prompts
Note
: CPR breathing coaching prompts can be set through the
Rescue options
menu option listed on the
AED
Main Menu
screen. The factory default setting provides prompts for chest compressions only (no breathing).
Breathing prompts can be included either by changing the menu option (refer to the
”
section in
Chapter 8 of this manual) or by pressing a softkey button during rescue. (Refer to
”
”
section of this chapter.)
Voice
Text
“Begin CPR now”
Begin CPR Now
Purpose:
This indicates that the user should begin performing CPR immediately. The unit will not monitor the
patient’s ECG rhythm during this CPR period.
“Give Compressions”
Give “xx” Compressions
Purpose:
This indicates that the user should begin CPR compressions immediately. The unit will emit a beep at
the rate that compressions should be given.
“Continue”
“Continue for 1 minute ‘xx’ seconds”
Continue “xx” Seconds
Purpose:
This indicates that the user should continue performing CPR. This phrase is spoken to let the user
know that the unit is still operating normally. The unit will not be monitoring the patient’s ECG rhythm during this
mandatory two minute CPR period. (AHA/ERC 2010 Protocol)
“Ending in 5, 4, 3, 2, 1”
Ending in “xx” Seconds
Purpose:
This indicates that the user should prepare to finish performing CPR. This phrase is spoken during the last
several seconds of the CPR period to let the operator know that the unit is still operating normally and that the CPR
period is ending.
“Stop CPR”
“Stop Now”
Stop CPR
Stop Now
Purpose:
This indicates that the CPR period has ended and the user should stop CPR. The unit will enter Analyze Mode.
“Do not touch the patient”
“Analyzing heart rhythm”
Do Not Touch Patient
Analyzing Rhythm
Purpose:
This indicates the unit has entered Analyze Mode and is performing an ECG analysis. The user should
not touch the patient during the ECG analysis.