Defibrillator – Manual Mode
(DDU-2400 only)
Category
Specification
Energy
User selectable:
25, 50, 70, 100, 150, 200 Joules (adult defibrillation pads)
25, 50, 70, 100, Joules (pediatric defibrillation pads)
Note:
energy is limited to 50 J only, when using DDP-2002
(attenuating pediatric defibrillation pads)
Charge control
User initiated, CHARGE softkey
Charge time
See battery charge time specifications in
of this manual
Charge complete indication
• SHOCK button flashing
• Two tone alarm
Shock delivery
Shock is delivered by a single SHOCK button
DISARM
Automatic
• Within 30 seconds after charge complete, if operator has not pressed
SHOCK button,
or
• If defibrillation pads are removed from patient or unplugged from unit.
Manual
• If operator presses the manual DISARM softkey or ON/OFF button pressed for
greater than 2 seconds
Defibrillator – ECG Mode
(DDU-2400 & DDU-2450 only)
Category
Specification
Displayed ECG
ECG information is received from defibrillation pads in anterior-lateral or
anterior-posterior positions, or from the 3-wire ECG Monitor Adapter in Lead II
Screen type
TFT Color LCD with backlight (53.6mm x 71.5mm) (320 x 240 pixels)
Displayed Range
Differential: +/- 2 mV full-scale
Sweep Speed
25 mm/sec
Frequency Response
1 Hz to 22 Hz (-3 dB), nominal
Sensitivity
10 mm/mV, nominal
Heart rate display
20 to 200 bpm, updated once/second
• Displayed when ECG viewing is enabled
• An out of range heart rate is indicated by dashed lines in the display
Note:
The ECG Display provides non-diagnostic ECG of the patient’s heart rhythm. It is not intended to provide
diagnostic or ST segment interpretation.
Note:
The ECG waveform sensitivity and magnification settings are fixed and are not user adjustable.
Defibtech DDU-2000 Series AED (continued)
11
.
Tec
hnical Specifications