Energy – Manual Mode
(DDP-2001 Adult and DDP-2003 Non-attenuating Pediatric Defibrillation Pads) (Nominal)
Load Impedance (Ohms)
Selected
Energy
25
50
75
100
125
150
175
25 J
25
25
26
26
26
25
25
50 J
50
50
51
51
51
50
49
70 J
70
70
71
71
72
70
68
100 J
99
100
101
101
102
100
97
150 J
153
151
152
151
153
146
142
200 J
194
195
198
197
201
195
189
DDP-2003 pediatric defibrillation pads Selected Energy range is limited from 25 to 100 J.
Phase Time and Energy
(DDP-2002 Attenuating Pediatric Defibrillation Pads)
Patient Impedance
(Ohms)
Phase A Duration
(ms)
Phase B Duration
(ms)
Energy Delivered
(Joules)
25
4.1
4.1
35
50
5.8
3.8
47
75
5.8
3.8
51
100
7.2
4.8
53
125
7.2
4.8
52
150
9
6
53
175
9
6
51
Applies to both AED and Manual Modes.
Note
: If impedance is out of range for proper analysis and shock delivery, voice and/or visual prompts will inform the user.
Defibtech DDU-2000 Series AED (continued)
11
.
Tec
hnical Specifications