8.3 Entering Maintenance Mode
Before You Begin:
Make certain the DDU-2000 Series AED is turned off and a battery pack is installed.
AED status
OK
Battery status
Expires 01/2017
Pads status
Adult
Expires 06/2017
AED S/N
099999999
Battery S/N
099999999
Software version
V2.1
STEP 1
– Press and release the CENTER softkey button.
Result –
The unit will turn on and display the AED Status
Screen for a short period of time.
If the unit does not turn on, check to make sure a good battery pack
is installed. (Refer to the
“
”
section in Chapter 7
of this manual
.
)
STEP 2 –
Press the
BOTTOM
softkey button
(to the right of the Tool Icon)
.
Note
:
If the bottom softkey button
is not pressed within a short period of time, the unit will automatically turn off.
Result –
The unit will enter Maintenance Mode and display the AED Main Menu screen.
8.4 AED Main Menu Screen
The AED Main Menu screen allows the user to view the status of the AED, perform maintenance functions, change
AED options, and access help topics. All maintenance functions are accessed through the AED Main Menu screen.
The user may select from the following options using the softkey buttons:
Rescue now
AED status
AED maintenance
AED options
Rescue options
Help topics
Turn AED off
AED Main Menu
Rescue now
– Puts device in Rescue Mode
AED status
– Displays current AED status information
AED maintenance
– Displays AED Maintenance Menu screen
AED options
– Displays AED Options Menu screen
Rescue options
– Displays Rescue Options Menu screen
Help topics
– Displays Help Topics screen
Turn AED off
– Turns unit off
When the user selects “Rescue now,” the unit will exit Maintenance Mode and proceed directly to Rescue Mode.
The other menu options will perform different functions and are described in detail below.
8.5 AED Status Screen
The AED Status screen displays unit-specific data such as current status, battery pack charge state, battery pack
expiration date, defibrillation pads expiration date, unit serial number, battery pack serial number, and software
version number.
Before you begin
: Be sure the unit is in Maintenance Mode.
To enter
: Navigate to
AED status
:
AED Main Menu
Ú
AED status
Note
: When the unit is turned off, the AED Status screen can also be accessed by pressing the center softkey button.
What it does
: The unit will display the AED Status screen. This is an informational screen only; no action is taken
by the AED.
To exit
: To exit the AED Status screen, press and release the BOTTOM softkey button.
The unit will exit the AED Status screen and return to the AED Main Menu screen.
8. Maintenance Mode
41
DAC-E2510EN-BA