Model 3770

 

DC STRAIN GAGE PANEL METER 

4.

CALIBRATION

This section contains the instructions for calibrating the 3770. Included is a functional description of the instrument 
front-panel (see Page 1). To perform calibration, proceed as follows.

( a )  

Connect Sensor and Analog terminals as required. Connect and apply power to the unit. The front-
panel digital display should light indicating the application of the AC input power. Allow 10 
minutes of warm up for stabilization of transducer characteristics. Remove the front cover of the 
3770 unit by removing the two small Phillips screws on the front panel. TARE should not be 
activated. 

( b )  

Center the Zero and Fine Span potentiometers by rotating the potentiometers CW until you feel the 
“end stop” or by the number of turns accomplished (Coarse is 22 turn, Fine is 25 turn). Rotate the 
potentiometer back to the middle of their range (Coarse 11 turns, Fine 12-13 turns). This will 
establish the adjustments middle authority. 

( c )  

Position the front panel switches to the desired settings for the application. Refer to Section 3 for 
details. 

( d )  

Relax the sensor input to “zero” load and adjust the Wide Zero and Fine Zero controls until the 
meter reading is achieved at the point which will be the “zero” value reference.  

( e )  

Once the Zero position has been established, apply a known load typically in the nominal working 
range of the sensor. With the load active and stabilized, adjust the Wide Span, Coarse Span and 
Fine Span controls until the display reading required is achieved.  

( f )  

Return the sensor to the "Zero" position and re-adjust the Coarse Zero and Fine Zero controls for 
the desired reading or analog output of the unit. Repeat the Span (step e) adjustment as necessary to 
achieve the proper reading or analog output. Note that changes in Span  (Gain)  will affect Zero. 
Span and Zero re-adjustments may need to be repeated to obtain the desired reading. (Note that the 
3770 meter also has independent display adjustment capability to where the digital display can be 
adjusted independently to the analog output signal of the unit so as to achieve proper analog output 
levels from the meter.) 

( g )  

Once the Zero and the Gain (SPAN) adjustments have been accomplished, if needed - the user can 
adjust for Negative non-symmetrical value (as referenced by the positive SPAN value) by utilizing 
the front panel Symmetry control when the load is Negative. To accomplish this, apply a known 
negative load and adjust the symmetry control for the correct readout or analog  output signal 
required. Note the Symmetry control has approx. ± 2 % full scale authority. 

( h )  

        Once completed, replace the front display lens cover to the original position and ensure proper shielding and 

grounding have been done to the meter. Minor calibration adjustment may be accomplished using the front 
panel potentiometer for zero and span. 

Display Calibration

 The 3770 instrument display has separate adjustment controls that are independent of the 

analog output controls - allowing the user to alter the display reading to a value that is suited for the display while 
maximizing the analog output amplitude. In the standard – default switch position (as marked near the Coarse 
Display Span control), the display has a fixed full scale reading set by the display range switches of 5000, 10000 or 
20000 counts. At theses settings the full scale analog output signal will be 5 Vdc or 10 Vdc, as selected via the 
analog controls. To accomplish the display adjustment change, the user would first set the desired display range 
and decimal location using the display dip switches, adjust the analog output to the desired level and then adjust the 
display Coarse then Fine span controls for the required digital readout. 

Page  1

1

 

Summary of Contents for 3770

Page 1: ...MODEL 3770 DC STRAIN GAGE CONDITIONER SB 1 1 INSTRUCTION MANUAL 3700 Instrument Series...

Page 2: ...of the analog full scale signal The Model 3770 can be either calibrated by the two point dead weight process involving known input parameters or through the Shunt method using front panel push buttons...

Page 3: ...30 ppm C vs Time 10 ppm month Linearity better than 0 03 of full scale Filter 3 pole modified Butterworth 3 dB down at 2 Hz 20 Hz or 200 Hz selectable Fast output always enabled 5 KHz response J3 Pin...

Page 4: ...el Function J1 1 N C No connection J1 2 N C No connection J1 3 HOLD Input Analog Hold command J1 4 TARE Input Analog TARE command J1 5 TARE ENABLE External TARE enable input J1 6 LOGIC COM Logic Commo...

Page 5: ...nd a shorting wire in place of the external shunt resistor position This provides the electrical path for the Calibration Sense line to enable the sensor s internal shunt resistor when the or Shunt Ca...

Page 6: ...rnal or external across the connected four arm bridge sensor for simulated calibration verification check of the sensor s electrical characteristics LOGIC COM R CAL N O Switch J2 HOLD TARE Connections...

Page 7: ...KHz Analog Voltage Output 5 Vdc or 10 Vdc with front panel selectable cutoff filter 2 20 200 Hz Analog Current Output 4 20mA 4 12 20mA at selected cutoff filter SIGNAL SIGNAL SIGNAL SIGNAL SIGNAL SIGN...

Page 8: ...selects Display Full Scale Range Range will equate to an analog output value of 5 Vdc or 10 Vdc per Analog Control switch placement Selection is set for 1 count in 5000 2 counts in 10000 and 5 counts...

Page 9: ...range gain for 0 5 to 10 mV V sensors Display Span Adjustments Used to adjust the digital readout of the 3770 meter the user calibrates the analog output for a specific span value and then adjusts th...

Page 10: ...ion voltage The full scale analog output will be 5 000 Vdc The current output will be set for 4 ma zero and 20 ma will positive full scale The analog low pass filter will be set for 20 Hz Shunt calibr...

Page 11: ...nse leads of the bridge CAL Negative Calibration Activates the SHUNT Negative by switching in the external or internal shunt calibration resistor across the Cal Sense and Sense leads of the bridge TAR...

Page 12: ...t Note that changes in Span Gain will affect Zero Span and Zero re adjustments may need to be repeated to obtain the desired reading Note that the 3770 meter also has independent display adjustment ca...

Page 13: ...potentiometers fully CW Coarse is 22 turns Fine is 25 turns then reverse direction CCW Coarse 11 turns and Fine 13 turns to obtain mid authority of the controls Typically this is done on initial cali...

Page 14: ...scale analog output signal will be 5 VDC or 10 VDC as selected by the Analog Controls of the unit To accomplish the display adjustment change the user would first set the desired display range and dec...

Page 15: ...a factor of 2 Wide Span Position SPAN Low mV V SPAN High mV V 0 3 90 5 80 1 3 30 4 90 2 2 90 4 20 3 2 50 3 60 4 2 10 3 10 5 1 80 2 70 6 1 50 2 30 7 1 30 1 90 8 1 10 1 70 9 1 00 1 40 A 0 84 1 20 B 0 72...

Page 16: ...hase price to the Buyer ANY PRODUCT FOUND TO BE DAMAGED THROUGH CUSTOMER NEGLIGENCE OR MISUSE MAY BE EXCLUDED FROM ANY AND ALL POLICIES CONTAINED IN THIS DOCUMENT ALL EQUIPMENT TO BE REPAIRED OR REPLA...

Reviews: