background image

Model 3770        

DC STRAIN GAGE PANEL METER 

4.

CALIBRATION – Shunt Calibration

This section contains the instructions for calibrating the 3770 using the Shunt calibration method.  

( a )  

Connect Power, Sensor and Analog terminals as required. Apply power. The front-panel digital 
display should light indicating the application of the AC input power. Allow 10 minutes of warm 
up for stabilization of transducer characteristics. Remove the front panel cover of the 3770 unit 
which is held in place by the two small Phillip head screws.  

Note: The 3770 unit has an internal 59K Shunt resistor installed within the unit. If a different 
resistor value is to be used to correspond to the transducer’s calibration sheet; install the resistor 
on the rear J1 terminals provided and place the Analog Control Switch 5 to External. Shunt 
calibration will only activate if the CAL Sense line is connected to the + or – Signal lines on the 
transducer input connector J2.  

For transducers with an internally installed shunt resistor:  1.) Cal Sense should be attached to the 
transducer’s shunt activation connector pin; 2.) External Shunt selected on the Analog Controls; 
3.) A shorting wire is to be installed in place of the External Shunt resistor terminals on the rear 
of the unit. 

( b )  

Center the Zero and Span potentiometers as needed by rotating the potentiometers fully CW   
(Coarse is 22 turns, Fine is 25 turns), then reverse direction - CCW (Coarse 11 turns and Fine 13 
turns) to obtain mid-authority of the controls. Typically this is done on initial calibration. When 
recalibrating for minor adjustments this may not be required. 

( c )  

  With mechanical Zero or Balance established at the transducer (transducer “relaxed”); Adjust 

Wide Zero then Coarse Zero and then Fine Zero for the Zero position reading on the display or 
the desired analog output signal. Display can be adjusted independently to the analog signal. 

( d )  

Depress and hold the + CAL button while adjusting the Wide Span, Coarse Span and Fine Span 
controls, as needed, for precise engineering units reading on the display as determined by the 
sensor’s calibration sheet or other reference data (see Equivalent Input shunt calculation below). 

( e )  

Repeat Steps (d) and (e) to obtain proper measurement readings since the Gain – Span controls 
will affect the Zero amplification of the input signal. 

( f )  

Once completed, if the sensor is going to be utilized in the Negative or CCW mode, depress and 
hold the – CAL button and adjust the Symmetry control for a proper symmetrical reading of the 
shunt input as noted on the sensor’s calibration data sheet. Note the Symmetry control affects the 
Negative Span for approx. ± 2% of full scale adjustment control. 

( g )  

Release the Cal button and adjust Zero as needed. Re-install the front panel to the instrument. 
Minor adjustments and controls are assessable via the front of the unit. 

Page  1

2

 

Summary of Contents for 3770

Page 1: ...MODEL 3770 DC STRAIN GAGE CONDITIONER SB 1 1 INSTRUCTION MANUAL 3700 Instrument Series...

Page 2: ...of the analog full scale signal The Model 3770 can be either calibrated by the two point dead weight process involving known input parameters or through the Shunt method using front panel push buttons...

Page 3: ...30 ppm C vs Time 10 ppm month Linearity better than 0 03 of full scale Filter 3 pole modified Butterworth 3 dB down at 2 Hz 20 Hz or 200 Hz selectable Fast output always enabled 5 KHz response J3 Pin...

Page 4: ...el Function J1 1 N C No connection J1 2 N C No connection J1 3 HOLD Input Analog Hold command J1 4 TARE Input Analog TARE command J1 5 TARE ENABLE External TARE enable input J1 6 LOGIC COM Logic Commo...

Page 5: ...nd a shorting wire in place of the external shunt resistor position This provides the electrical path for the Calibration Sense line to enable the sensor s internal shunt resistor when the or Shunt Ca...

Page 6: ...rnal or external across the connected four arm bridge sensor for simulated calibration verification check of the sensor s electrical characteristics LOGIC COM R CAL N O Switch J2 HOLD TARE Connections...

Page 7: ...KHz Analog Voltage Output 5 Vdc or 10 Vdc with front panel selectable cutoff filter 2 20 200 Hz Analog Current Output 4 20mA 4 12 20mA at selected cutoff filter SIGNAL SIGNAL SIGNAL SIGNAL SIGNAL SIGN...

Page 8: ...selects Display Full Scale Range Range will equate to an analog output value of 5 Vdc or 10 Vdc per Analog Control switch placement Selection is set for 1 count in 5000 2 counts in 10000 and 5 counts...

Page 9: ...range gain for 0 5 to 10 mV V sensors Display Span Adjustments Used to adjust the digital readout of the 3770 meter the user calibrates the analog output for a specific span value and then adjusts th...

Page 10: ...ion voltage The full scale analog output will be 5 000 Vdc The current output will be set for 4 ma zero and 20 ma will positive full scale The analog low pass filter will be set for 20 Hz Shunt calibr...

Page 11: ...nse leads of the bridge CAL Negative Calibration Activates the SHUNT Negative by switching in the external or internal shunt calibration resistor across the Cal Sense and Sense leads of the bridge TAR...

Page 12: ...t Note that changes in Span Gain will affect Zero Span and Zero re adjustments may need to be repeated to obtain the desired reading Note that the 3770 meter also has independent display adjustment ca...

Page 13: ...potentiometers fully CW Coarse is 22 turns Fine is 25 turns then reverse direction CCW Coarse 11 turns and Fine 13 turns to obtain mid authority of the controls Typically this is done on initial cali...

Page 14: ...scale analog output signal will be 5 VDC or 10 VDC as selected by the Analog Controls of the unit To accomplish the display adjustment change the user would first set the desired display range and dec...

Page 15: ...a factor of 2 Wide Span Position SPAN Low mV V SPAN High mV V 0 3 90 5 80 1 3 30 4 90 2 2 90 4 20 3 2 50 3 60 4 2 10 3 10 5 1 80 2 70 6 1 50 2 30 7 1 30 1 90 8 1 10 1 70 9 1 00 1 40 A 0 84 1 20 B 0 72...

Page 16: ...hase price to the Buyer ANY PRODUCT FOUND TO BE DAMAGED THROUGH CUSTOMER NEGLIGENCE OR MISUSE MAY BE EXCLUDED FROM ANY AND ALL POLICIES CONTAINED IN THIS DOCUMENT ALL EQUIPMENT TO BE REPAIRED OR REPLA...

Reviews: