Model 3770        

DC STRAIN GAGE PANEL METER 

2

C

O N N E C T I O N S

Fig. 6  - Transducer Cabling – DC Strain Gage- J2

 

+ SENSE

- SENSE

+ EXCITATION

- EXCITATION

+ SIGNAL

SHIELD

- SIGNAL

CAL SENSE

 EXCITATION

 EXCITATION

 SENSE

  SENSE

 SIGNAL

  SIGNAL

CAL SENSE

+

_

_

_

+

+

SHIELD

Note:

 

Sense lines must be connected at the transducer or at the instrument for proper operation. 

J3 - Analog Connections 

 

Provides analog outputs from the meter's signal conditioning area in the form 

of ±5Vdc or ±  10 Vdc (selectable via the front panel controls) and 4-20 mA (or selectable 4-12-20 mA). 
Full scale output is referenced to the display "range" selected (5000, 10000, or 20000). I.e. If range 
10000 is selected, the analog output will be 5 Vdc or 10 Vdc, as determined by the front panel controls 
(or 20 mA FS) when in the default/home position and the display reads 10000 (irrespective of the 
decimal point). 

1 2 3 4 5

Analog Voltage

Output

+/- 5 Vdc or +/- 10 Vdc with 

fixed cutoff filter @ 5 KHz

Analog Voltage

Output

+/- 5 Vdc or +/- 10 Vdc with  

front panel selectable cutoff 

filter @ 2/20/200 Hz

Analog Current

Output

4-20mA, 4-12-20mA

at selected cutoff filter 

+ SIGNAL

- SIGNAL

- SIGNAL

+ SIGNAL

+ SIGNAL

- SIGNAL

J3 -  Analog Output Connections 

Pin 1  

Voltage Common (- Signal out) reference for Pin 2 and Pin 3 + Signal Voltage Output. 

Pin 2  

Voltage Output Selected (+Signal out) reference to Pin 1. The Full Scale Voltage is determined by 
the position of switch 

1 -VDC

 on the Conditioner Controls. Output response is determined by the 

position of switch 

3-FIL

 and 

4- FIL 

on the Analog Controls. 

Pin 3

  

Voltage Output Fixed. (+ Signal out) reference to Pin 1. Full Scale Voltage is determined by the 
position of switch 

1-VDC

 of the Conditioner Controls. Filter response is fixed at the highest 

analog signal response of 5 KHz. 

Pin 4

  

4-20 mA Current Output Common (- Signal) reference for Pin 5 Current Output. 

Pin 5

  

4-20 mA Current Output (+Signal) reference to Pin 4 Current Output Common. The Current 
Output is selectable for 4 - 20 mA or 4-12-20 mA as determined by the position of switch 

2 -mA

 

of the Conditioner Control switch(s) on the Front Panel. 

Page 

6

 

Summary of Contents for 3770

Page 1: ...MODEL 3770 DC STRAIN GAGE CONDITIONER SB 1 1 INSTRUCTION MANUAL 3700 Instrument Series...

Page 2: ...of the analog full scale signal The Model 3770 can be either calibrated by the two point dead weight process involving known input parameters or through the Shunt method using front panel push buttons...

Page 3: ...30 ppm C vs Time 10 ppm month Linearity better than 0 03 of full scale Filter 3 pole modified Butterworth 3 dB down at 2 Hz 20 Hz or 200 Hz selectable Fast output always enabled 5 KHz response J3 Pin...

Page 4: ...el Function J1 1 N C No connection J1 2 N C No connection J1 3 HOLD Input Analog Hold command J1 4 TARE Input Analog TARE command J1 5 TARE ENABLE External TARE enable input J1 6 LOGIC COM Logic Commo...

Page 5: ...nd a shorting wire in place of the external shunt resistor position This provides the electrical path for the Calibration Sense line to enable the sensor s internal shunt resistor when the or Shunt Ca...

Page 6: ...rnal or external across the connected four arm bridge sensor for simulated calibration verification check of the sensor s electrical characteristics LOGIC COM R CAL N O Switch J2 HOLD TARE Connections...

Page 7: ...KHz Analog Voltage Output 5 Vdc or 10 Vdc with front panel selectable cutoff filter 2 20 200 Hz Analog Current Output 4 20mA 4 12 20mA at selected cutoff filter SIGNAL SIGNAL SIGNAL SIGNAL SIGNAL SIGN...

Page 8: ...selects Display Full Scale Range Range will equate to an analog output value of 5 Vdc or 10 Vdc per Analog Control switch placement Selection is set for 1 count in 5000 2 counts in 10000 and 5 counts...

Page 9: ...range gain for 0 5 to 10 mV V sensors Display Span Adjustments Used to adjust the digital readout of the 3770 meter the user calibrates the analog output for a specific span value and then adjusts th...

Page 10: ...ion voltage The full scale analog output will be 5 000 Vdc The current output will be set for 4 ma zero and 20 ma will positive full scale The analog low pass filter will be set for 20 Hz Shunt calibr...

Page 11: ...nse leads of the bridge CAL Negative Calibration Activates the SHUNT Negative by switching in the external or internal shunt calibration resistor across the Cal Sense and Sense leads of the bridge TAR...

Page 12: ...t Note that changes in Span Gain will affect Zero Span and Zero re adjustments may need to be repeated to obtain the desired reading Note that the 3770 meter also has independent display adjustment ca...

Page 13: ...potentiometers fully CW Coarse is 22 turns Fine is 25 turns then reverse direction CCW Coarse 11 turns and Fine 13 turns to obtain mid authority of the controls Typically this is done on initial cali...

Page 14: ...scale analog output signal will be 5 VDC or 10 VDC as selected by the Analog Controls of the unit To accomplish the display adjustment change the user would first set the desired display range and dec...

Page 15: ...a factor of 2 Wide Span Position SPAN Low mV V SPAN High mV V 0 3 90 5 80 1 3 30 4 90 2 2 90 4 20 3 2 50 3 60 4 2 10 3 10 5 1 80 2 70 6 1 50 2 30 7 1 30 1 90 8 1 10 1 70 9 1 00 1 40 A 0 84 1 20 B 0 72...

Page 16: ...hase price to the Buyer ANY PRODUCT FOUND TO BE DAMAGED THROUGH CUSTOMER NEGLIGENCE OR MISUSE MAY BE EXCLUDED FROM ANY AND ALL POLICIES CONTAINED IN THIS DOCUMENT ALL EQUIPMENT TO BE REPAIRED OR REPLA...

Reviews: