 DC STRAIN GAGE PANEL METER 

Tech Tip on use of wide range settings for the DC Strain Gage Conditioner 

Due to multiple amplifier stages within the 3770 instrument, attention to the proper gain setting and understanding of 
the sensor inputs should be reviewed to produce a linear- amplified analog output signal and display reading. 

The wide span – gain control (as shown in the front panel diagram) has 16 positions to amplify the incoming signal by 
incremental steps. Each step is positioned so the Coarse and Fine Span potentiometer controls overlap each step to 
provide a continuous linear gain of the signal from 0.5 mV/V to 10.00 mV/V full scale. Below is a table of each wide 
gain steps, indicating the nominal low and high input signal range per position – full scale in mV/V to the meter.     

Note: When using 5 V excitation, increase the table values by a factor of 2.

 

Wide Span 

Position 

SPAN-Low 

 mV / V 

SPAN-High 

 mV / V 

3.90 

5.80 

3.30 

4.90 

2.90 

4.20 

2.50 

3.60 

2.10 

3.10 

1.80 

2.70 

1.50 

2.30 

1.30 

1.90 

1.10 

1.70 

1.00 

1.40 

0.84 

1.20 

0.72 

1.00 

0.62 

0.92 

0.53 

0.80 

0.45 

0.65 

0.38 

0.58 

DC Strain Gage Sensors sensor: 

DC Strain Gage Sensors will typically specify their electrical sensitivity in mV/V for traditional foil gage transducers 
or in mV full scale for semiconductor type sensors; either type will function with the 3770 unit as dictated by the 
excitation voltage needed along with the full scale output of the sensor. The 3770 has selectable excitation of 5 and 
10 Vdc and a maximum specified input of 50.0 mV (with 20% over-range) into the meter. As a reference, the above 
table can be used as a guide to determine the working full scale range of the transducer during calibration and the 
proper position of the Wide Span control. Note that additional gain or less gain can be accomplished as needed if an 
expanded or reduced area of the sensor measurement level is required. 

Page 1

4

 

Summary of Contents for 3770

Page 1: ...MODEL 3770 DC STRAIN GAGE CONDITIONER SB 1 1 INSTRUCTION MANUAL 3700 Instrument Series...

Page 2: ...of the analog full scale signal The Model 3770 can be either calibrated by the two point dead weight process involving known input parameters or through the Shunt method using front panel push buttons...

Page 3: ...30 ppm C vs Time 10 ppm month Linearity better than 0 03 of full scale Filter 3 pole modified Butterworth 3 dB down at 2 Hz 20 Hz or 200 Hz selectable Fast output always enabled 5 KHz response J3 Pin...

Page 4: ...el Function J1 1 N C No connection J1 2 N C No connection J1 3 HOLD Input Analog Hold command J1 4 TARE Input Analog TARE command J1 5 TARE ENABLE External TARE enable input J1 6 LOGIC COM Logic Commo...

Page 5: ...nd a shorting wire in place of the external shunt resistor position This provides the electrical path for the Calibration Sense line to enable the sensor s internal shunt resistor when the or Shunt Ca...

Page 6: ...rnal or external across the connected four arm bridge sensor for simulated calibration verification check of the sensor s electrical characteristics LOGIC COM R CAL N O Switch J2 HOLD TARE Connections...

Page 7: ...KHz Analog Voltage Output 5 Vdc or 10 Vdc with front panel selectable cutoff filter 2 20 200 Hz Analog Current Output 4 20mA 4 12 20mA at selected cutoff filter SIGNAL SIGNAL SIGNAL SIGNAL SIGNAL SIGN...

Page 8: ...selects Display Full Scale Range Range will equate to an analog output value of 5 Vdc or 10 Vdc per Analog Control switch placement Selection is set for 1 count in 5000 2 counts in 10000 and 5 counts...

Page 9: ...range gain for 0 5 to 10 mV V sensors Display Span Adjustments Used to adjust the digital readout of the 3770 meter the user calibrates the analog output for a specific span value and then adjusts th...

Page 10: ...ion voltage The full scale analog output will be 5 000 Vdc The current output will be set for 4 ma zero and 20 ma will positive full scale The analog low pass filter will be set for 20 Hz Shunt calibr...

Page 11: ...nse leads of the bridge CAL Negative Calibration Activates the SHUNT Negative by switching in the external or internal shunt calibration resistor across the Cal Sense and Sense leads of the bridge TAR...

Page 12: ...t Note that changes in Span Gain will affect Zero Span and Zero re adjustments may need to be repeated to obtain the desired reading Note that the 3770 meter also has independent display adjustment ca...

Page 13: ...potentiometers fully CW Coarse is 22 turns Fine is 25 turns then reverse direction CCW Coarse 11 turns and Fine 13 turns to obtain mid authority of the controls Typically this is done on initial cali...

Page 14: ...scale analog output signal will be 5 VDC or 10 VDC as selected by the Analog Controls of the unit To accomplish the display adjustment change the user would first set the desired display range and dec...

Page 15: ...a factor of 2 Wide Span Position SPAN Low mV V SPAN High mV V 0 3 90 5 80 1 3 30 4 90 2 2 90 4 20 3 2 50 3 60 4 2 10 3 10 5 1 80 2 70 6 1 50 2 30 7 1 30 1 90 8 1 10 1 70 9 1 00 1 40 A 0 84 1 20 B 0 72...