Model 3178
R and C BALANCE Controls:
These controls are used to set the output
to zero when the transducer is unloaded.
NULL Pushbutton:
This pushbutton is pressed when nulling the trans-
ducer bridge with the R and C BALANCE controls. It provides for
non-synchronous demodulation of the carrier for balancing purposes.
After a minimum reading is obtained on the display device by adjusting
the C control, the pushbutton is released and the output is zeroed using
the R controls.
SPAN Controls:
The
Coarse
and
Fine
SPAN controls are used to set the
output to the
dead weight
value when
dead weight
calibration is used.
They can also be used to set the output to the
Equivalent Input
value
when the CAL (+ or –) button is pressed.
CAL Pushbuttons:
The + and – CAL pushbuttons provide for shunt
calibration in the positive and negative realms, respectively. They are
used in conjunction with the SPAN controls to calibrate the instrument.
When both positive and negative realm calibration is required, the
+CAL button is used in conjunction with the SPAN controls and the
–CAL button is used with the internal
Symmetry
adjustment.
Figure 5. Front-Panel Description
12
Summary of Contents for 3000 series
Page 1: ...MODEL 3178 STRAIN GAGE CONDITIONER SB 5 1 INSTRUCTION MANUAL 3000 Instrument Series...
Page 6: ...This page intentionally blank...
Page 12: ...Daytronic Corporation Figure 3 Instrument Panel Mounting 5...
Page 16: ...I600 Fig 4 I O Wiring Data See the CORRECTION to Fig 4 in the front of this manual...
Page 24: ...This page intentionally blank...
Page 25: ...Fig 6 Block Diagram...
Page 29: ...This page intentionally blank...
Page 30: ...Daytronic Corporation Dayton OH 800 668 4745 www daytronic com...