Daytronic Corporation
values. Solder two resistors together, then solder the remaining two
resistors together. Next, connect the two junctions together using a
separate wire as shown. There is a good reason for this method of
construction, and it should be followed. Connect the substitute or simu-
lated transducer to the instrument I/O connector using a short 4-wire
cable configuration as shown in Figure 4. Attempt to balance the substi-
tute or simulated transducer. If conditions now appear to be normal, the
transducer or cable is at fault. If the previous difficulties persist, the 3178
is faulty.
21
Summary of Contents for 3000 series
Page 1: ...MODEL 3178 STRAIN GAGE CONDITIONER SB 5 1 INSTRUCTION MANUAL 3000 Instrument Series...
Page 6: ...This page intentionally blank...
Page 12: ...Daytronic Corporation Figure 3 Instrument Panel Mounting 5...
Page 16: ...I600 Fig 4 I O Wiring Data See the CORRECTION to Fig 4 in the front of this manual...
Page 24: ...This page intentionally blank...
Page 25: ...Fig 6 Block Diagram...
Page 29: ...This page intentionally blank...
Page 30: ...Daytronic Corporation Dayton OH 800 668 4745 www daytronic com...