background image

Model 3178

cable  provides  connection  from  the  3178  calibration  circuit  to  the  calibration

resistor  which  is  integral  to  the 

1600 Series 

transducers,  without  requiring  the

removal  of  the  3178  calibration  resistor.

When  user-fabricated  transducer  cabling  is  used,  it  should  take  the  form  of

either  the  4-,  5-,  or  7-wire  cable  configuration  shown  in  Figure  4.  The  4-wire

configuration  should  be  used  when  overall 

deadweight 

calibration  is  the  method

used  and  the  required  cable  length  is  less  than  20  feet.  The  5-wire  configuration

should  be  used  when  the  instrument  is  to  be  calibrated  by  achieving  a  precisely

known 

Equivalent Input 

value  through  the  use  of  a  shunt  calibration  resistor

supplied by the transducer manufacturer and when the required cable length is less

than 20 feet. The 7-wire configuration should be used with cable lengths longer than

20  feet  since  the  excitation  voltage  is  sensed  and  regulated  at  the  transducer  and

optimum  shunt  calibration  can  be  achieved.

Remote Calibration Check. 

 The instrument can be placed in the calibration

mode (positive realm only) by shorting pins 5 (Signal Common) and 8 (Remote Cal)

of  the  rear-panel  I/O  connector.  Figure  4  indicates  three  methods  of  remotely

entering  the  calibration  mode  (external  switch,  transistor,  or  TTL  source).  The

Remote Cal 

function  provides  a  convenient  method  of  periodically  monitoring

calibration  of  the  instrument  in  the  positive  realm.

Master/Slave Connections. 

 When more than one 3178 (or a combination of

3178 or 3130 LVDT Conditioners) is being used in a measurement setup (instru-

ments are continuously mounted or the transducer cabling is in a common conduit

or raceway), beat frequencies may be produced from the 3-kHz oscillators used in

the instruments to develop the excitation. To prevent beat frequencies from occur-

ring, one unit can be designated the 

master, 

and the remaining units can be driven

from the oscillator contained in the 

master 

unit. The remaining units are designated

as slave instruments. To perform 

master/slave 

wiring, refer to Figure 4.

Analog Outputs. 

  Two  analog  outputs  are  available  at  the  instrument  I/O

connector, with each output having a different passband: dc to 2 Hz and dc to 400

Hz.  The  cutoff  frequencies  are  achieved  with  active  low-pass  filters.  When  the

dc-to-2  Hz  output  is  used,  a  trade  off  is  made  between  noise  elimination  and

increased  time-to-answer  or  slew  time.  Each  output  has  a  60-dB rolloff a decade

from  the  cutoff  frequency.  The  filter  characteristics  are  given  by  the  following

equations.

8

Summary of Contents for 3000 series

Page 1: ...MODEL 3178 STRAIN GAGE CONDITIONER SB 5 1 INSTRUCTION MANUAL 3000 Instrument Series...

Page 2: ...reprinted reproduced or used in any form or by any electronic mechanical or other means including photocopying and recording or in any information storage and retrieval system without permission in wr...

Page 3: ...aired with CAL SENSE unconnected at Connector A SHIELD Daytronic 400 Series Transducer Connections 1 A 2 B 3 4 C 9 D A B C EXC EXC SENSE SENSE SIG SIG CAL SENSE B A Extra Wire paired with CAL SENSE un...

Page 4: ...ction 4 Wire Configuration for Cables Shorter Than 20 Feet SENSE EXC SENSE SIG SIG CAL SENSE 1 A 2 B 3 C 4 9 A SHIELD Extra Wire paired with CAL SENSE unconnected at Connector A EXC 6 Wire Configurati...

Page 5: ...DEL 3178 STRAIN GAGE CONDITIONER INSTRUCTION MANUAL Model 3178 Instruction Manual v SB 5 1 Pub No 3178M 5 1 Issued 03 01 Part No 91121 Daytronic Corporation Dayton OH Tel 800 668 4745 www daytronic co...

Page 6: ...This page intentionally blank...

Page 7: ...onditioner 1 2 Instrument Mounting Dimensions 4 3 Instrument Panel Mounting 5 4 I O Wiring 9 5 Front Panel Description 12 6 Block Diagram 17 7 Star Bridge Construction 20 LIST OF TABLES Table Page 1 S...

Page 8: ...ssary balancing and calibration controls and conditions amplifies the applied input to a standard Five Volt Data Signal Level which is the output analog signal level of 3000 Series Instruments Two ana...

Page 9: ...e Analog Outputs Two analog outputs available 0 to 5 volts with 50 overrange 5 milliamperes maximum Bandpass is dc to 2 Hz or dc to 400 Hz Active low pass filters provide for rolloff of 60 dB per deca...

Page 10: ...cteristics Positive realm calibra tion can be remotely checked by means of Remote Cal terminals on the instrument I O connector The 3178 Strain Gage Conditioner is also available in two additional for...

Page 11: ...Model 3178 A Bench Mounting B Rack Mounting C Panel Mounting Figure 2 Instrument Mounting Dimensions 4...

Page 12: ...Daytronic Corporation Figure 3 Instrument Panel Mounting 5...

Page 13: ...the front panel screwdriver adjustments and buttons are accessible These screws give approxi mately 1 8 inch of adjustment consequently this is the maximum panel thickness which should be used CAUTION...

Page 14: ...d even though the transducer calibration data mentions some other resistance value In Section 3 of this manual the tech niques described above are demonstrated If however the installed value of calibr...

Page 15: ...tor Figure 4 indicates three methods of remotely entering the calibration mode external switch transistor or TTL source The Remote Cal function provides a convenient method of periodically monitoring...

Page 16: ...I600 Fig 4 I O Wiring Data See the CORRECTION to Fig 4 in the front of this manual...

Page 17: ...I600 I600 I600 I600 I600 I600 I600 Fig 4 cont d See the CORRECTION to Fig 4 in the front of this manual...

Page 18: ...ALANCE control to obtain the minimum least positive or most nega tive output reading In some instances an integral digital indicator will be used to display the conditioner output Model 3278 or 3378 W...

Page 19: ...rols SPAN Controls The Coarse and Fine SPAN controls are used to set the output to the dead weight value when dead weight calibration is used They can also be used to set the output to the Equivalent...

Page 20: ...h the transducer loaded as in step f adjust the Symmetry control just to the right of the CAL button to obtain a dead weight reading equal to that obtained in step e If dead weight calibration is not...

Page 21: ...ide switch is used to turn ON primary power Overload protection is provided by a 0 25 ampere fuse mounted near the ac connection point When the slide switch is ON primary power is applied to the power...

Page 22: ...the output of the Integrat ing Amplifier changes the resistance of an Automatic Gain Control element FET to return the oscillator amplitude to its nominal value of 2 volts ac The Integrating Amplifier...

Page 23: ...ontacts of the CAL switch Calibration in the positive realm can also be remotely checked When the Remote Cal input at the 3178 I O connector is brought to a zero volt ground level through the action o...

Page 24: ...This page intentionally blank...

Page 25: ...Fig 6 Block Diagram...

Page 26: ...B within a decade of the cutoff frequency 5 VERIFICATION OF NORMAL OPERATION It is the purpose of this section to aid the user in determining in the event of a malfunction to which the Model 3178 is s...

Page 27: ...a transducer and cable known to be in good condition or by simulating a balanced transducer using either a commercially available transducer simulator or the simple star bridge arrangement shown in F...

Page 28: ...ason for this method of construction and it should be followed Connect the substitute or simu lated transducer to the instrument I O connector using a short 4 wire cable configuration as shown in Figu...

Page 29: ...This page intentionally blank...

Page 30: ...Daytronic Corporation Dayton OH 800 668 4745 www daytronic com...

Reviews: