Model 3178
from the
+Signal
terminal of the transducer to the + or –
Excitation
terminals. The
Excitation
terminal to which the connection is made is
determined by the direction of the loading or off-zero reading.
(c) The inability to balance correctly where the instrument output reads
totally off scale and the BALANCE controls have no authority can very
likely be the result of a damaged or defective transducer or cable. This
possibility can be confirmed (or eliminated) by substituting a transducer
and cable known to be in good condition or by simulating a balanced
transducer, using either a commercially available transducer simulator
or the simple
star bridge
arrangement shown in Figure 7. The
star bridge
simulates a conventional four-arm bridge in an exact condition of bal-
ance. To construct a
star bridge
connect four 10% carbon resistors as
shown in Figure 7. Use 180-ohm resistors to simulate a 350-ohm bridge
and use 56-ohm resistors to simulate a 120-ohm bridge. Neither the
resistor values nor temperature characteristics are critical since the
balance condition of a
star bridge
is not determined by the resistance
Figure 7. Star Bridge Construction
20
Summary of Contents for 3000 series
Page 1: ...MODEL 3178 STRAIN GAGE CONDITIONER SB 5 1 INSTRUCTION MANUAL 3000 Instrument Series...
Page 6: ...This page intentionally blank...
Page 12: ...Daytronic Corporation Figure 3 Instrument Panel Mounting 5...
Page 16: ...I600 Fig 4 I O Wiring Data See the CORRECTION to Fig 4 in the front of this manual...
Page 24: ...This page intentionally blank...
Page 25: ...Fig 6 Block Diagram...
Page 29: ...This page intentionally blank...
Page 30: ...Daytronic Corporation Dayton OH 800 668 4745 www daytronic com...